Safety profile, electrical performances and clinical outcomes of a novel leadless pacemaker; first results from real-world clinical practice in Saudi Arabia

Abstract Introduction Leadless pacemakers (LP) are emerging as an effective and safe alternative to transvenous pacemakers to reduce lead-related and pocket-related complications and to improve patient experience. A new LP system, the Aveir VR, has been approved in 2022 for US and other countries, and in 2023 for Europe, and features innovative characteristics such as the active fixation helix mechanism and the possibility to measure sensing/pacing parameters before LP deployment. Purpose Evaluate safety, electrical performances and clinical outcomes associated with the first Aveir VR LP implants in 2 Hospitals in Saudi Arabia. Methods We performed a prospective evaluation of all-comer patients who received Aveir VR LP. Decision for LP implant was based on conditions such as cardiac pacing indication, lack of upper extremity venous access, risk of device infection, and life expectancy, and using shared decision-making. Clinical and electrical data were collected at implant and at 1st and 6th month visits. Results A total of 54 patients (25 female, mean age 69 years) were included between December 2022 and October 2023. Pacemaker implant indication was AV block in 25 (46%) patients, sinus node disease in 19 (35%) patients, atrial fibrillation and slow ventricular response in 9 (17%) patients and left bundle branch block in 1 (2%) patient. Implant was successful in all attempted cases. Final device position was lower septum in 51/54 (94.4%) patients and mid-septum in 3/54 (5.6%) patients. Mean procedure time was 48±22 minutes and showed a decreasing trend as a function of operator experience with a mean of 55, 50 and 36 minutes in the first, second and third group of consecutive implants, respectively. Mean fluoroscopic time was 11±8 minutes and showed a decreasing trend as a function of operator experience with a mean of 13, 13 and 6 minutes in the first, second and third group of consecutive implants, respectively. Final device position was achieved at first attempt in 42/54 (77.8%) cases and at first or second attempt in 51/54 (94.4%) cases. At implant mean R wave sensing was 7.6±3.7 mV, mean impedance was 724±335 Ohm, mean threshold was 1.0±0.6 V at 0.4 ms. During implant there was one patient with pericardial effusion treated with pericardiocentesis. In a mean follow-up of 73 days there were no procedure-related or device-related complications, electrical parameters showed acceptable values, and all implanted devices performed as expected. Median longevity, estimated at last follow-up visit, was 19 years (interquartile range: 12-20 years). Conclusion First real-world cases of use of a novel leadless pacemaker in Saudi Arabia show optimal feasibility and good safety profile and electrical performances. Procedure and fluoroscopic times were acceptable and decreased as a function of operator experience. Longevity estimations predict very long service life-times.

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.707
Citations
1
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Safety profile, electrical performances and clinical outcomes of a novel leadless pacemaker; first results from real-world clinical practice in Saudi Arabia

Andrea Grammatico, Raed Sweidan, FaisalM Al-Samadi, Ali Chami et al.
1 citations
European Heart Journal
Cardiac pacing and defibrillation studies
0.40
article

Safety profile, electrical performances and clinical outcomes of a novel leadless pacemaker; first results from real-world clinical practice in Saudi Arabia

Andrea Grammatico, Raed Sweidan, FaisalM Al-Samadi, Ali Chami, FayezA Bokhari
article en
1 citations

Abstract

Abstract Introduction Leadless pacemakers (LP) are emerging as an effective and safe alternative to transvenous pacemakers to reduce lead-related and pocket-related complications and to improve patient experience. A new LP system, the Aveir VR, has been approved in 2022 for US and other countries, and in 2023 for Europe, and features innovative characteristics such as the active fixation helix mechanism and the possibility to measure sensing/pacing parameters before LP deployment. Purpose Evaluate safety, electrical performances and clinical outcomes associated with the first Aveir VR LP implants in 2 Hospitals in Saudi Arabia. Methods We performed a prospective evaluation of all-comer patients who received Aveir VR LP. Decision for LP implant was based on conditions such as cardiac pacing indication, lack of upper extremity venous access, risk of device infection, and life expectancy, and using shared decision-making. Clinical and electrical data were collected at implant and at 1st and 6th month visits. Results A total of 54 patients (25 female, mean age 69 years) were included between December 2022 and October 2023. Pacemaker implant indication was AV block in 25 (46%) patients, sinus node disease in 19 (35%) patients, atrial fibrillation and slow ventricular response in 9 (17%) patients and left bundle branch block in 1 (2%) patient. Implant was successful in all attempted cases. Final device position was lower septum in 51/54 (94.4%) patients and mid-septum in 3/54 (5.6%) patients. Mean procedure time was 48±22 minutes and showed a decreasing trend as a function of operator experience with a mean of 55, 50 and 36 minutes in the first, second and third group of consecutive implants, respectively. Mean fluoroscopic time was 11±8 minutes and showed a decreasing trend as a function of operator experience with a mean of 13, 13 and 6 minutes in the first, second and third group of consecutive implants, respectively. Final device position was achieved at first attempt in 42/54 (77.8%) cases and at first or second attempt in 51/54 (94.4%) cases. At implant mean R wave sensing was 7.6±3.7 mV, mean impedance was 724±335 Ohm, mean threshold was 1.0±0.6 V at 0.4 ms. During implant there was one patient with pericardial effusion treated with pericardiocentesis. In a mean follow-up of 73 days there were no procedure-related or device-related complications, electrical parameters showed acceptable values, and all implanted devices performed as expected. Median longevity, estimated at last follow-up visit, was 19 years (interquartile range: 12-20 years). Conclusion First real-world cases of use of a novel leadless pacemaker in Saudi Arabia show optimal feasibility and good safety profile and electrical performances. Procedure and fluoroscopic times were acceptable and decreased as a function of operator experience. Longevity estimations predict very long service life-times.

European Heart JournalVol. 45(Supplement_1)
Abbott (Sweden) (SE), King Fahd Medical City (SA), King Fahd Armed Forces Hospital (SA), Abbott (Germany) (DE)
Openalex Percentile: Top 31%
Cardiac pacing and defibrillation studies
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.