Infection management in the extravascular implantable cardioverter defibrillator: experience from the EV ICD Pivotal study

Abstract Background The extravascular implantable cardioverter defibrillator (EV ICD) has recently emerged as an option for guideline indicated ICD patients. Results from the EV ICD Pilot and Pivotal studies demonstrated that the device can provide safe and effective therapy through chronic follow up. Lead placement outside the vasculature has the potential to reduce systemic infections. Infection occurrence and management details from the Pivotal are reported. Purpose To report on details of infection occurrence and management from the EV ICD Pivotal study. Methods The EV ICD Pivotal study was an international, prospective, single-arm, premarket clinical study. Patients with a class I or IIa indication for a single-chamber ICD per ESC or ACC/AHA/HRS guidelines were enrolled. Results In the EV ICD Pivotal study, 316 patients had an implant attempt, (74.7% male, age 53.8±13.1 years, 82% primary prevention, LVEF of 38.9%±15.4%, and NYHA Class I [23.7%] or II/III [65.5%]), of which 15 had a system or procedure-related infection (4.7%, 9 - 737 days post-implant) through an average 16.2 months follow-up. Of these infections, 8 were classified as major complications, 3 as minor, and 4 as observations. Most (n=9) infections were addressed through medication with or without wound care (2.8% of patients) while 6 (1.9% of patients) resulted in system removal. There was a single lead-related infection addressed through system explant. However, there were no reports of mediastinitis, sepsis, or endocarditis. No system removal resulted in sequelae. Conclusions Infection rates observed in the EV ICD Pivotal study were similar to those reported in the subcutaneous implantable cardioverter defibrillator. The results from this study support that EV ICD related infections are treatable with antibiotic therapy or system removal. No system or procedure-related systemic infections such as mediastinitis, sepsis, or endocarditis have been reported in the Pivotal study.

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.746
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Infection management in the extravascular implantable cardioverter defibrillator: experience from the EV ICD Pivotal study

Nicolas Clémenty, Bradley P. Knight, Paul A. Friedman, Christopher Wiggenhorn et al.
European Heart Journal
Cardiac pacing and defibrillation studies
article

Infection management in the extravascular implantable cardioverter defibrillator: experience from the EV ICD Pivotal study

Nicolas Clémenty, Bradley P. Knight, Paul A. Friedman, Christopher Wiggenhorn, Christophe Leclercq, Ian Crozier, Ulrika Birgersdotter‐Green, Darius P. Sholevar, Thomas A. Mattioni, Jaimie Manlucu, Francis Murgatroyd, Ahmad Hersi, Lucas V.A. Boersma
article en

Abstract

Abstract Background The extravascular implantable cardioverter defibrillator (EV ICD) has recently emerged as an option for guideline indicated ICD patients. Results from the EV ICD Pilot and Pivotal studies demonstrated that the device can provide safe and effective therapy through chronic follow up. Lead placement outside the vasculature has the potential to reduce systemic infections. Infection occurrence and management details from the Pivotal are reported. Purpose To report on details of infection occurrence and management from the EV ICD Pivotal study. Methods The EV ICD Pivotal study was an international, prospective, single-arm, premarket clinical study. Patients with a class I or IIa indication for a single-chamber ICD per ESC or ACC/AHA/HRS guidelines were enrolled. Results In the EV ICD Pivotal study, 316 patients had an implant attempt, (74.7% male, age 53.8±13.1 years, 82% primary prevention, LVEF of 38.9%±15.4%, and NYHA Class I [23.7%] or II/III [65.5%]), of which 15 had a system or procedure-related infection (4.7%, 9 - 737 days post-implant) through an average 16.2 months follow-up. Of these infections, 8 were classified as major complications, 3 as minor, and 4 as observations. Most (n=9) infections were addressed through medication with or without wound care (2.8% of patients) while 6 (1.9% of patients) resulted in system removal. There was a single lead-related infection addressed through system explant. However, there were no reports of mediastinitis, sepsis, or endocarditis. No system removal resulted in sequelae. Conclusions Infection rates observed in the EV ICD Pivotal study were similar to those reported in the subcutaneous implantable cardioverter defibrillator. The results from this study support that EV ICD related infections are treatable with antibiotic therapy or system removal. No system or procedure-related systemic infections such as mediastinitis, sepsis, or endocarditis have been reported in the Pivotal study.

European Heart JournalVol. 45(Supplement_1)
Mayo Clinic (US), King's College London (GB), London Health Sciences Centre (CA), Christchurch Hospital (NZ), University Medical Center (US), Medtronic (Ireland) (IE), University of California San Diego (US), Northwestern University (PH), Clinique du Millénaire (FR), St. Antonius Ziekenhuis (NL), Hôpital Pontchaillou (FR), HonorHealth (US), King Khalid University Hospital (SA), King Khalid University (SA)
Openalex Percentile: Top 66%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.