Automated cardiac arrest detection using a photoplethysmography wristband: algorithm performance in relation to sex, skin color type and arm hair density

Abstract Background While survival from witnessed out-of-hospital cardiac arrest (OHCA) has improved, this is not the case for unwitnessed cardiac arrest as the event goes unnoticed. Automated cardiac arrest detection and alerting through a wristband is considered a promising technique also facilitating early assistance for this large group. In the DETECT-1 study, a photoplethysmography(PPG) algorithm for circulatory arrest detection was developed in patients with induced circulatory arrest. However, patient characteristics, such as skin color and arm hair density, can influence the accuracy of the PPG signal. Thus, we performed further analyses to study the effect of patient characteristics on the algorithm’s performance. Purpose To assess the performance of the developed PPG algorithm for cardiac arrest detection in relation to sex, skin color type and arm hair density. Methods Patients who underwent short-lasting circulatory arrest induction during transcatheter aortic valve implantation, ventricular tachycardia ablation or subcutaneous ICD implantation wore a PPG wristband during the procedure. Continuous ECG and invasive blood pressure recordings served as reference standards. A PPG algorithm for circulatory arrest detection was developed. The algorithm’s performance was assessed in subgroups based on sex, skin color type (according to the Fitzpatrick scale), and arm hair density. The sensitivity for the detection of circulatory arrest and false positive alarms were compared between the different subgroups. Results In total, 151 patients were included in the current analysis. Analyzed data consisted of 159 hours of PPG data including 176 induced circulatory arrest events. The population consisted of 57 (38%) females and 94 (62%) males. The sensitivity for circulatory arrest detection was 100% in the female group vs. 98% in the male group (p=ns); false positive alarms occurred in 4 females and 9 males, resulting in a positive predictive value (PPV) of 94% and 92% (p=ns), respectively. Within the skin color subgroup analysis, 114 (75%) patients had skin color type I or II, and 37 (25%) patients had a skin color type III, IV, V or VI. There were no differences in sensitivity (p=ns) and PPV (p=ns) in the skin color subgroup (table 1). The population consisted of 106 (71%) patients with nil or sparse arm hair density, 43 (29%) patients with moderate or dense arm hair density, and 2 not reported. There were no differences in sensitivity (p=ns) and PPV (p=ns) within the arm hair density subgroup (table 1). Conclusions Automated cardiac arrest detection using a smartwatch is feasible and its accuracy is not affected by sex, skin color type or arm hair density in this study. Further study is needed to assess performance of the algorithm in a larger sample size and a more diverse study population. If proven to be effective, automated cardiac arrest detection has the potential to improve survival from OHCA.Sensitivity/PPV for arrest detection

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.3544
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated cardiac arrest detection using a photoplethysmography wristband: algorithm performance in relation to sex, skin color type and arm hair density

Nicolas M. Van Mieghem, Aysun Cetinyurek‐Yavuz, Eelko Ronner, Niels T B Scholte et al.
European Heart Journal
Non-Invasive Vital Sign Monitoring
article

Automated cardiac arrest detection using a photoplethysmography wristband: algorithm performance in relation to sex, skin color type and arm hair density

Nicolas M. Van Mieghem, Aysun Cetinyurek‐Yavuz, Eelko Ronner, Niels T B Scholte, Ross Edgar, Masih Mafi-Rad, Sing‐Chien Yap, Marc A. Brouwer, Kevin Vernooy, Niels van Royen, Judith Bonnes, Kambiz Ebrahimkheil, Eric Boersma, R J Beukema, Peter Stas
article en

Abstract

Abstract Background While survival from witnessed out-of-hospital cardiac arrest (OHCA) has improved, this is not the case for unwitnessed cardiac arrest as the event goes unnoticed. Automated cardiac arrest detection and alerting through a wristband is considered a promising technique also facilitating early assistance for this large group. In the DETECT-1 study, a photoplethysmography(PPG) algorithm for circulatory arrest detection was developed in patients with induced circulatory arrest. However, patient characteristics, such as skin color and arm hair density, can influence the accuracy of the PPG signal. Thus, we performed further analyses to study the effect of patient characteristics on the algorithm’s performance. Purpose To assess the performance of the developed PPG algorithm for cardiac arrest detection in relation to sex, skin color type and arm hair density. Methods Patients who underwent short-lasting circulatory arrest induction during transcatheter aortic valve implantation, ventricular tachycardia ablation or subcutaneous ICD implantation wore a PPG wristband during the procedure. Continuous ECG and invasive blood pressure recordings served as reference standards. A PPG algorithm for circulatory arrest detection was developed. The algorithm’s performance was assessed in subgroups based on sex, skin color type (according to the Fitzpatrick scale), and arm hair density. The sensitivity for the detection of circulatory arrest and false positive alarms were compared between the different subgroups. Results In total, 151 patients were included in the current analysis. Analyzed data consisted of 159 hours of PPG data including 176 induced circulatory arrest events. The population consisted of 57 (38%) females and 94 (62%) males. The sensitivity for circulatory arrest detection was 100% in the female group vs. 98% in the male group (p=ns); false positive alarms occurred in 4 females and 9 males, resulting in a positive predictive value (PPV) of 94% and 92% (p=ns), respectively. Within the skin color subgroup analysis, 114 (75%) patients had skin color type I or II, and 37 (25%) patients had a skin color type III, IV, V or VI. There were no differences in sensitivity (p=ns) and PPV (p=ns) in the skin color subgroup (table 1). The population consisted of 106 (71%) patients with nil or sparse arm hair density, 43 (29%) patients with moderate or dense arm hair density, and 2 not reported. There were no differences in sensitivity (p=ns) and PPV (p=ns) within the arm hair density subgroup (table 1). Conclusions Automated cardiac arrest detection using a smartwatch is feasible and its accuracy is not affected by sex, skin color type or arm hair density in this study. Further study is needed to assess performance of the algorithm in a larger sample size and a more diverse study population. If proven to be effective, automated cardiac arrest detection has the potential to improve survival from OHCA.Sensitivity/PPV for arrest detection

European Heart JournalVol. 45(Supplement_1)
Radboud University Nijmegen (NL), Maastricht University Medical Centre (NL), Radboud University Medical Center (NL), Reinier de Graaf Hospital (NL), Erasmus University Rotterdam (NL)
Good health and well-being
Openalex Percentile: Top 83%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.