"Honey pot"-like lesion formation: impact of catheter contact angle on lesion formation by novel temperature controlled ablation catheter in a porcine experimental model

Abstract Background The DiamondTemp(DT) catheter enables precise temperature-controlled ablation. However, the effects of the DT catheter's contact angle on lesion formation are poorly understood. Purpose This study evaluated lesion formation using the DT catheter at different angles in a porcine experimental model. Methods We used freshly sacrificed porcine hearts, and radiofrequency catheter ablation with DT was performed at 50 watts for 15 seconds at an upper-temperature setting of 60°C. The contact force (5, 10, and 30 grams) and catheter contact angles (30°, 45°, and 90°) were changed in each set (n = 13 each). The surface width (SW), maximum lesion width (MW), lesion depth (LD), surface area (SA), and distance from the distal edge to the widest area (DW) were evaluated. In addition, the generator impedance drop during radiofrequency application was analyzed. Results Figure 1 shows the summarized data in this study. SW and MW were longer at 30° than at 90° (SW 5 g: 6.34 ± 0.91 vs. 5.07 ± 0.87 mm; SW 10 g: 7.46 ± 0.55. vs. 5.60 ± 0.84 mm; SW 30 g: 7.42 ± 0.74. vs. 6.14 ± 0.77 mm; MW 10 g: 8.81 ± 0.55. vs. 7.89 ± 0.84 mm; MW 30 g: 9.24 ± 0.67. vs. 8.03 ± 1.04 mm, p <0.05, respectively). There were no significant differences in LD by catheter angle except at 30 g. SA was larger at 30° than at 90° (SA 5 g: 30.03 ± 0.91. vs. 23.61 ± 0.87; SA 10 g: 29.24 ± 0.55. vs. 21.27 ± 0.84; SA 30 g: 35.86 ± 0.74. vs. 31.74 ± 0.77, p <0.05, respectively). Besides, DW was longer at 30° than at 90° (DW 5 g: 4.28 ± 1.17. vs. 2.50 ± 0.32 mm; DW 10 g: 4.94 ± 0.52. vs. 2.47 ± 0.38 mm; DW 30 g: 5.34 ± 0.26. vs. 2.96 ± 0.40 mm, p <0.05, respectively). There were no significant differences in the impedance drop according to the catheter angle. Conclusion With the DT catheters, the lesion width increased at a shallower contact angle, whereas the LD did not. SA also increased at a shallower contact angle. The DT catheter created a large ablation lesion on the proximal side of the catheter, which looked like a "honey pot." (Figure 2)summarized dataImage of honey pot-like lesion formation

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.385
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

"Honey pot"-like lesion formation: impact of catheter contact angle on lesion formation by novel temperature controlled ablation catheter in a porcine experimental model

Jun Kishihara, Y. Masuda, Yusuke Murayama, Hidehira Fukaya et al.
European Heart Journal
Cardiac Arrhythmias and Treatments
article

"Honey pot"-like lesion formation: impact of catheter contact angle on lesion formation by novel temperature controlled ablation catheter in a porcine experimental model

Jun Kishihara, Y. Masuda, Yusuke Murayama, Hidehira Fukaya, Jun Oikawa, Shinichi Niwano, Y Yazaki, Emiyu Ogawa, Shuhei Kobayashi, Junya Ako, Daiki Saito, Hironori Nakamura, Naruya Ishizue, Hitoshi Mori, Seiji OGISO
article en

Abstract

Abstract Background The DiamondTemp(DT) catheter enables precise temperature-controlled ablation. However, the effects of the DT catheter's contact angle on lesion formation are poorly understood. Purpose This study evaluated lesion formation using the DT catheter at different angles in a porcine experimental model. Methods We used freshly sacrificed porcine hearts, and radiofrequency catheter ablation with DT was performed at 50 watts for 15 seconds at an upper-temperature setting of 60°C. The contact force (5, 10, and 30 grams) and catheter contact angles (30°, 45°, and 90°) were changed in each set (n = 13 each). The surface width (SW), maximum lesion width (MW), lesion depth (LD), surface area (SA), and distance from the distal edge to the widest area (DW) were evaluated. In addition, the generator impedance drop during radiofrequency application was analyzed. Results Figure 1 shows the summarized data in this study. SW and MW were longer at 30° than at 90° (SW 5 g: 6.34 ± 0.91 vs. 5.07 ± 0.87 mm; SW 10 g: 7.46 ± 0.55. vs. 5.60 ± 0.84 mm; SW 30 g: 7.42 ± 0.74. vs. 6.14 ± 0.77 mm; MW 10 g: 8.81 ± 0.55. vs. 7.89 ± 0.84 mm; MW 30 g: 9.24 ± 0.67. vs. 8.03 ± 1.04 mm, p <0.05, respectively). There were no significant differences in LD by catheter angle except at 30 g. SA was larger at 30° than at 90° (SA 5 g: 30.03 ± 0.91. vs. 23.61 ± 0.87; SA 10 g: 29.24 ± 0.55. vs. 21.27 ± 0.84; SA 30 g: 35.86 ± 0.74. vs. 31.74 ± 0.77, p <0.05, respectively). Besides, DW was longer at 30° than at 90° (DW 5 g: 4.28 ± 1.17. vs. 2.50 ± 0.32 mm; DW 10 g: 4.94 ± 0.52. vs. 2.47 ± 0.38 mm; DW 30 g: 5.34 ± 0.26. vs. 2.96 ± 0.40 mm, p <0.05, respectively). There were no significant differences in the impedance drop according to the catheter angle. Conclusion With the DT catheters, the lesion width increased at a shallower contact angle, whereas the LD did not. SA also increased at a shallower contact angle. The DT catheter created a large ablation lesion on the proximal side of the catheter, which looked like a "honey pot." (Figure 2)summarized dataImage of honey pot-like lesion formation

European Heart JournalVol. 45(Supplement_1)
Kitasato University (JP), Saitama Medical University (JP)
Openalex Percentile: Top 66%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.