Safety and efficacy of leadless pacemaker extraction using a novel retrieval tool: a multicenter analysis

Abstract Background Leadless pacemakers (LP) have emerged as a safe and effective alternative for cardiac pacing, increasingly adopted by the medical community. While rare, complications such as pacemaker syndrome or device dislodgment can occur. In such instances, the extraction of the leadless pacemaker and its replacement with a new device should be considered. Purpose This multicenter study aimed to describe the use of the Aveir retrieval sheath during Micra extraction procedure and to assess the success rates and potential complications associated with the procedure. Methods In this observational study, we analyzed prospectively collected data from 14 patients that underwent leadless pacemaker using Abbott retrieval catheter at three high volume centers for pacemaker extraction and leadless implant in United States. Thought femoral vein the Aveir steerable retrieval catheter is advanced to the right atrium and then across the tricuspid valve. Under TEE and fluoroscopy guidance the snare is advanced over the Micra, and the button is secured. A protective sleeve is advanced over the LP. With application of traction to snare and countertraction with the sleeve against the endocardium, the LP is free from the myocardium and removed via the sheath. (Fig. 1) The procedural duration should be intended as the combined total of the time taken for the extraction of the existing pacemaker and the implantation time of the new device. Extractions performed during the first year were described as early and thereafter as late. Procedural success was described as complete extraction of LP without procedural complication. Results The reason for early explant was LP dislodgment and for late explant was pacemaker syndrome. Median time to extraction was 4±2 months in early explant and 35±9 months in late explant. Procedural time in early and late groups was 68.7 and 72.3 min respectively (P=0.32). Fluoroscopy time in early and late groups was 12.2 and 16.4 respectively (P= 0.18). LP was then replaced with a transvenous pacemaker. Success was 100% in both groups. No complication occurred in both groups. Conclusion Micra pacemaker using Aveir retrieval catheter is an effective tool irrespective of dwell time.

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.749
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Safety and efficacy of leadless pacemaker extraction using a novel retrieval tool: a multicenter analysis

Luigi Di Biase, Nicola Pierucci, Rodney Horton, Carola Gianni et al.
European Heart Journal
Cardiac pacing and defibrillation studies
article

Safety and efficacy of leadless pacemaker extraction using a novel retrieval tool: a multicenter analysis

Luigi Di Biase, Nicola Pierucci, Rodney Horton, Carola Gianni, Devi Nair, Weeranun D. Bode, Brandon Doty, Domenico Giovanni Della Rocca, Prem Geeta Torlapati, V La Fazia, Kirollos Gabrah, Mohanty Sanghamitra, Amin Al‐Ahmad, Andrea Natale, Mohamad Bassiouny
article en

Abstract

Abstract Background Leadless pacemakers (LP) have emerged as a safe and effective alternative for cardiac pacing, increasingly adopted by the medical community. While rare, complications such as pacemaker syndrome or device dislodgment can occur. In such instances, the extraction of the leadless pacemaker and its replacement with a new device should be considered. Purpose This multicenter study aimed to describe the use of the Aveir retrieval sheath during Micra extraction procedure and to assess the success rates and potential complications associated with the procedure. Methods In this observational study, we analyzed prospectively collected data from 14 patients that underwent leadless pacemaker using Abbott retrieval catheter at three high volume centers for pacemaker extraction and leadless implant in United States. Thought femoral vein the Aveir steerable retrieval catheter is advanced to the right atrium and then across the tricuspid valve. Under TEE and fluoroscopy guidance the snare is advanced over the Micra, and the button is secured. A protective sleeve is advanced over the LP. With application of traction to snare and countertraction with the sleeve against the endocardium, the LP is free from the myocardium and removed via the sheath. (Fig. 1) The procedural duration should be intended as the combined total of the time taken for the extraction of the existing pacemaker and the implantation time of the new device. Extractions performed during the first year were described as early and thereafter as late. Procedural success was described as complete extraction of LP without procedural complication. Results The reason for early explant was LP dislodgment and for late explant was pacemaker syndrome. Median time to extraction was 4±2 months in early explant and 35±9 months in late explant. Procedural time in early and late groups was 68.7 and 72.3 min respectively (P=0.32). Fluoroscopy time in early and late groups was 12.2 and 16.4 respectively (P= 0.18). LP was then replaced with a transvenous pacemaker. Success was 100% in both groups. No complication occurred in both groups. Conclusion Micra pacemaker using Aveir retrieval catheter is an effective tool irrespective of dwell time.

European Heart JournalVol. 45(Supplement_1)
Policlinico Umberto I (IT), St Bernard's Hospital (GI), Texas Cardiac Arrhythmia (US), Rhythm (United states) (US)
Good health and well-being
Openalex Percentile: Top 66%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.