Accuracy of 3 Intraoral Scanners in Recording Impressions for Full Arch Dental Implant-Supported Prosthesis: An In Vitro Study

BACKGROUND This study used an edentulous mandibular resin model with 6 parallel osteotomy sites and aimed to compare the accuracy (trueness and precision) of 10 digital impressions using 3 intraoral scanners, the 3Shape TRIOS 5, Medit i700, and Primescan, using Medit Link v3.3.2 software. MATERIAL AND METHODS A model simulating a patient's lower jaw was surgically prepared at 6 parallel sites (implant osteotomy), allowing placement of 6 implant analogues. Matrix-Direct transfer abutments were attached to the analogs, and a reference scan was obtained using a CeramilMap 600 extraoral scanner. Three intraoral scanners (3Shape TRIOS 5, Medit i700, and Primescan) made 10 digital impressions of each model. The data obtained were superimposed and compared using software (Medit Link 3.3.2) to evaluate accuracy. Mean values were statistically analyzed using one-way ANOVA and post hoc Tukey test. Differences were considered significant at a P value of less than 0.05. RESULTS The TRIOS 5 intraoral scanner displayed the lowest deviation for precision (37.8±4.53 μm) and trueness (54.9±11 μm), followed by Medit i700 (precision 40.6±4.17 μm, trueness 60.5±10.9 μm), whereas the highest deviation (precision: 49.1±8.31 μm, trueness: 72.3±10.4 μm) was reported when Primescan intraoral scanner was used for recording impressions of full arch implants. When the 3 intraoral scanners were compared, a statistically significant difference was observed in terms of precision (P<0.005) and trueness (P<0.005). CONCLUSIONS TRIOS 5 intraoral scanner displayed the lowest deviation values for precision and trueness (more accurate), followed by Medit i700 and Primescan intraoral scanners. However, deviation values of all scanners were within clinically acceptable limits.

Authors

Institutions

Publication Details

Journal
Medical Science Monitor
Published
2024-10-28
DOI
https://doi.org/10.12659/msm.946624
Citations
9
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
4.44
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Accuracy of 3 Intraoral Scanners in Recording Impressions for Full Arch Dental Implant-Supported Prosthesis: An In Vitro Study

Mohammed E. Sayed, Khurshid Mattoo, Majed S. Altoman, Ahid Amer Alshahrani et al.
9 citations
Medical Science Monitor
Dental Implant Techniques and Outcomes
4.44
article

Accuracy of 3 Intraoral Scanners in Recording Impressions for Full Arch Dental Implant-Supported Prosthesis: An In Vitro Study

Mohammed E. Sayed, Khurshid Mattoo, Majed S. Altoman, Ahid Amer Alshahrani, Saad Saleh AlResayes, Saurabh Jain, Hossam F. Jokhadar, Saeed M. Alqahtani, Mohammad Alamri, Abdullah Hasan Alshehri, Hind Ziyad Al-najjar, Reem Abdullah A. Khawaji, Ghada Ali J. Hakami, Eman Hassan M. Solan, Manal A. Daish
article en
9 citations

Abstract

BACKGROUND This study used an edentulous mandibular resin model with 6 parallel osteotomy sites and aimed to compare the accuracy (trueness and precision) of 10 digital impressions using 3 intraoral scanners, the 3Shape TRIOS 5, Medit i700, and Primescan, using Medit Link v3.3.2 software. MATERIAL AND METHODS A model simulating a patient's lower jaw was surgically prepared at 6 parallel sites (implant osteotomy), allowing placement of 6 implant analogues. Matrix-Direct transfer abutments were attached to the analogs, and a reference scan was obtained using a CeramilMap 600 extraoral scanner. Three intraoral scanners (3Shape TRIOS 5, Medit i700, and Primescan) made 10 digital impressions of each model. The data obtained were superimposed and compared using software (Medit Link 3.3.2) to evaluate accuracy. Mean values were statistically analyzed using one-way ANOVA and post hoc Tukey test. Differences were considered significant at a P value of less than 0.05. RESULTS The TRIOS 5 intraoral scanner displayed the lowest deviation for precision (37.8±4.53 μm) and trueness (54.9±11 μm), followed by Medit i700 (precision 40.6±4.17 μm, trueness 60.5±10.9 μm), whereas the highest deviation (precision: 49.1±8.31 μm, trueness: 72.3±10.4 μm) was reported when Primescan intraoral scanner was used for recording impressions of full arch implants. When the 3 intraoral scanners were compared, a statistically significant difference was observed in terms of precision (P<0.005) and trueness (P<0.005). CONCLUSIONS TRIOS 5 intraoral scanner displayed the lowest deviation values for precision and trueness (more accurate), followed by Medit i700 and Primescan intraoral scanners. However, deviation values of all scanners were within clinically acceptable limits.

Medical Science MonitorVol. 30
King Abdulaziz University (SA), King Saud University (SA), King Khalid University (SA), Jazan University (SA)
Openalex Percentile: Top 5%
Dental Implant Techniques and Outcomes
4.44
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.