Efficacy and safety of the Aria pulmonary endovascular device in pulmonary hypertension

Abstract Background and aims A common feature of various forms of pulmonary hypertension (PH) is progressive decline of pulmonary arterial compliance (CPA), which correlates with reduced survival. In this acute study, we evaluated feasibility, safety and hemodynamic performance of the Aria pulmonary endovascular device in patients with PH associated with left heart disease (PH-LHD) and chronic lung disease (PH-CLD). Methods 8 patients with PH-LHD and 10 patients with PH-CLD were included in this study. The device was placed in the main pulmonary artery via the right femoral vein and was connected by a catheter to a gas-filled reservoir outside the body. During systole, gas shifts from the balloon to the reservoir, leading to deflation of the balloon. In diastole, the gas returns from the reservoir to the balloon, leading to balloon inflation and enhancing diastolic blood flow to the distal pulmonary capillary bed. Hemodynamics were assessed at baseline, and again with device off, device on and device off. The primary safety endpoint was the incidence of serious adverse events through 30 days after the procedure. Results No complications or investigational device-related serious adverse events occurred. Device activation in PH-LHD and PH-CLD patients decreased pulmonary arterial pulse pressure by 5.6±4.2mmHg (-12%; p=0.003) and 4.2±2.2mmHg (-11%; p<0.001), increased CPA by 0.4±0.2mL/mmHg (+23%; p=0.004) and 0.4±0.3mL/mmHg (+25%; p=0.001), and increased RV-PV coupling by 0.24±0.18 (+40%; p=0.012) and 0.11±0.07 (+21%; p=0.001), respectively. Conclusions Temporary implantation of the Aria endovascular device was feasible and safe. Device activation resulted in acute improvement of CPA and RV-PV coupling.Figure

Authors

Institutions

Publication Details

Journal
European Heart Journal
Published
2024-10-01
DOI
https://doi.org/10.1093/eurheartj/ehae666.2188
Citations
1
Primary Topic
Pulmonary Hypertension Research and Treatments
Type
article
Field-Weighted Citation Impact
0.42
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Efficacy and safety of the Aria pulmonary endovascular device in pulmonary hypertension

Christian Gerges, V Hartig, Nika Skoro‐Sajer, Varius Dannenberg et al.
1 citations
European Heart Journal
Pulmonary Hypertension Research and Treatments
0.42
article

Efficacy and safety of the Aria pulmonary endovascular device in pulmonary hypertension

Christian Gerges, V Hartig, Nika Skoro‐Sajer, Varius Dannenberg, Marc Pritzker, Iréne Lang, K Vollmer, John Scandurra, Ε. Kenneth Weir
article en
1 citations

Abstract

Abstract Background and aims A common feature of various forms of pulmonary hypertension (PH) is progressive decline of pulmonary arterial compliance (CPA), which correlates with reduced survival. In this acute study, we evaluated feasibility, safety and hemodynamic performance of the Aria pulmonary endovascular device in patients with PH associated with left heart disease (PH-LHD) and chronic lung disease (PH-CLD). Methods 8 patients with PH-LHD and 10 patients with PH-CLD were included in this study. The device was placed in the main pulmonary artery via the right femoral vein and was connected by a catheter to a gas-filled reservoir outside the body. During systole, gas shifts from the balloon to the reservoir, leading to deflation of the balloon. In diastole, the gas returns from the reservoir to the balloon, leading to balloon inflation and enhancing diastolic blood flow to the distal pulmonary capillary bed. Hemodynamics were assessed at baseline, and again with device off, device on and device off. The primary safety endpoint was the incidence of serious adverse events through 30 days after the procedure. Results No complications or investigational device-related serious adverse events occurred. Device activation in PH-LHD and PH-CLD patients decreased pulmonary arterial pulse pressure by 5.6±4.2mmHg (-12%; p=0.003) and 4.2±2.2mmHg (-11%; p<0.001), increased CPA by 0.4±0.2mL/mmHg (+23%; p=0.004) and 0.4±0.3mL/mmHg (+25%; p=0.001), and increased RV-PV coupling by 0.24±0.18 (+40%; p=0.012) and 0.11±0.07 (+21%; p=0.001), respectively. Conclusions Temporary implantation of the Aria endovascular device was feasible and safe. Device activation resulted in acute improvement of CPA and RV-PV coupling.Figure

European Heart JournalVol. 45(Supplement_1)
University of Minnesota Medical Center (US), Medical University of Vienna (AT)
Openalex Percentile: Top 31%
Pulmonary Hypertension Research and Treatments
0.42
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.