Evaluation of extra‐corporeal membrane oxygenator cannulae in pulsatile and non‐pulsatile pediatric mock circuits
Under non-pulsatile conditions, the system was characterized by overall higher pressures, resulting in higher support to the patient. The consequent increase of potential energy compensates for increases of kinetic energy, leading to a higher total hemodynamic energy. Pressure gradients and M number are independent of the testing conditions. Pulsatile testing conditions led to more physiological testing conditions, and it is recommended for ECMO testing.