Hemoglobin Tacoma is known to potentially interfere HbA1c assays. The variant is common in Finland with prevalence of up to 2% regionally and cases are also reported in areas that have attracted Finnish immigrants, especially in Sweden and North America. Here, we investigated the effect of Hb Tacoma on seven HbA1c methods. 20 non-variant and 20 Hb Tacoma samples were measured with Tina-quant Gen. 3 (immunoassay, considered as reference) and the following point of care instruments: Afinion 2, HbA1c 501 (both utilizing boronate affinity), QuikRead go, cobas b 101, DCA Atellica, and Standard F (all immunoassays). Repeatability was also assessed by measuring both non-variant and Hb Tacoma samples five times each at two different levels. For non-variant samples, the mean relative bias with all methods was < ±4%, whereas for Hb Tacoma samples Standard F had 38% mean relative bias. In absolute bias, the difference was 17 mmol/mol on average and constant through the measured range. For other methods the mean relative bias for Hb Tacoma samples was < ±6%. The repeatability with all methods was similar for non-variant and Hb Tacoma samples and at highest 4.1% (mean CV% of two levels). The observed interference by Standard F is likely due to two-antibody assay design as Hb Tacoma has been shown to result in conformational change. This interference is clinically significant and highlight the need for better controlling and better understanding hemoglobin variants in HbA1c testing.

Authors

Institutions

Publication Details

Journal
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2024-10-27
DOI
https://doi.org/10.1080/00365513.2024.2417383
Citations
1
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Hb Tacoma by seven HbA1c methods – one with significant interference

Päivikki Kangastupa, Anni Mäenpää, Moona Kangastie
1 citations
Scandinavian Journal of Clinical and Laboratory Investigation
Cardiac Arrhythmias and Treatments
0.40
article

Hb Tacoma by seven HbA1c methods – one with significant interference

Päivikki Kangastupa, Anni Mäenpää, Moona Kangastie
article en
1 citations

Abstract

Hemoglobin Tacoma is known to potentially interfere HbA1c assays. The variant is common in Finland with prevalence of up to 2% regionally and cases are also reported in areas that have attracted Finnish immigrants, especially in Sweden and North America. Here, we investigated the effect of Hb Tacoma on seven HbA1c methods. 20 non-variant and 20 Hb Tacoma samples were measured with Tina-quant Gen. 3 (immunoassay, considered as reference) and the following point of care instruments: Afinion 2, HbA1c 501 (both utilizing boronate affinity), QuikRead go, cobas b 101, DCA Atellica, and Standard F (all immunoassays). Repeatability was also assessed by measuring both non-variant and Hb Tacoma samples five times each at two different levels. For non-variant samples, the mean relative bias with all methods was < ±4%, whereas for Hb Tacoma samples Standard F had 38% mean relative bias. In absolute bias, the difference was 17 mmol/mol on average and constant through the measured range. For other methods the mean relative bias for Hb Tacoma samples was < ±6%. The repeatability with all methods was similar for non-variant and Hb Tacoma samples and at highest 4.1% (mean CV% of two levels). The observed interference by Standard F is likely due to two-antibody assay design as Hb Tacoma has been shown to result in conformational change. This interference is clinically significant and highlight the need for better controlling and better understanding hemoglobin variants in HbA1c testing.

Scandinavian Journal of Clinical and Laboratory InvestigationVol. 84(7-8)
Tampere University of Applied Sciences (FI), Seinäjoki University of Applied Sciences (FI), Tampere University (FI)
Openalex Percentile: Top 31%
Cardiac Arrhythmias and Treatments
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.