Comparative Study of Cerebrospinal Fluid Evaluation in Patients With Acute Leukemia Using the Manual Method and Sysmex XN-1000

Background Many pathological settings can be provided by the cytological analysis of cerebrospinal fluid (CSF). In our work, we aimed to evaluate the concordance between the analysis using the reference manual method and the analysis using Sysmex XN-1000<sup>®</sup>. Methods In our study, we examined 121 CSF samples. To count cells in different samples, we used two methods: a manual method using Fast Read 102<sup>®</sup> counting chambers performed by at least two experienced hematologists and Sysmex XN-1000<sup>®</sup>. Evaluation of the interchangeability and concordance between the two methods was possible thanks to Passing-Bablok regression and Bland-Altman graphs. Results Our study showed no evidence of concordance between the manual method and the automated method using Sysmex XN-1000<sup>®</sup> in cell counting on CSF samples. Conclusion Every clinical laboratory aims to use automated and standardized analysis methods in counting and differentiating cells, reducing response time and interlaboratory variability. Our study shows an absence of evidence of concordance between the manual and automated methods using Sysmex XN-1000 in cell counting on CSF samples.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2024-10-26
DOI
https://doi.org/10.7759/cureus.72423
Primary Topic
Intensive Care Unit Cognitive Disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Study of Cerebrospinal Fluid Evaluation in Patients With Acute Leukemia Using the Manual Method and Sysmex XN-1000

Rachid Seddik, M. Bensalah, Hasnae Lekfif, Abdelilah Berhili et al.
Cureus
Intensive Care Unit Cognitive Disorders
article

Comparative Study of Cerebrospinal Fluid Evaluation in Patients With Acute Leukemia Using the Manual Method and Sysmex XN-1000

Rachid Seddik, M. Bensalah, Hasnae Lekfif, Abdelilah Berhili, Assya Khermach
article en

Abstract

Background Many pathological settings can be provided by the cytological analysis of cerebrospinal fluid (CSF). In our work, we aimed to evaluate the concordance between the analysis using the reference manual method and the analysis using Sysmex XN-1000<sup>®</sup>. Methods In our study, we examined 121 CSF samples. To count cells in different samples, we used two methods: a manual method using Fast Read 102<sup>®</sup> counting chambers performed by at least two experienced hematologists and Sysmex XN-1000<sup>®</sup>. Evaluation of the interchangeability and concordance between the two methods was possible thanks to Passing-Bablok regression and Bland-Altman graphs. Results Our study showed no evidence of concordance between the manual method and the automated method using Sysmex XN-1000<sup>®</sup> in cell counting on CSF samples. Conclusion Every clinical laboratory aims to use automated and standardized analysis methods in counting and differentiating cells, reducing response time and interlaboratory variability. Our study shows an absence of evidence of concordance between the manual and automated methods using Sysmex XN-1000 in cell counting on CSF samples.

CureusVol. 16(10)
Centre Hospitalier Universitaire Mohammed VI (MA), Mohamed I University (MA)
Good health and well-being
Openalex Percentile: Top 76%
Intensive Care Unit Cognitive Disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.