The impact of defibrillation lead position on electrical parameters measured perioperatively and in long-term follow-up

<ns3:p>Introduction: Implantable cardioverter-defibrillators (ICD) with transvenous defibrillation leads remain the leading therapeutic method in the treatment of patients at high risk of sudden cardiac death. The lack of clear data confirming the superiority of any position in the right ventricle for placement of the defibrillation lead results in significant individualization of practice between operators.Aims: The study aims to assess the electrical parameters of defibrillation leads obtained intraoperatively and in the long-term follow-up, depending on the position of the lead in the right ventricle.Methods: Between 2018 and 2021, consecutive patients were enrolled in an observational study after the implantation of a device with defibrillation capability. Electrical parameters were measured during each control visit, 3 months after the implantation procedure, and then every 6-months during follow-up visits. The data regarding intraprocedural course (procedure time, an X-ray time, intraprocedural electrical parameters) were collected retrospectively. The patients were divided into two groups depending on the placement of the defibrillation lead in the right ventricle: the RVA (right ventricular apex group) and the non-RVA group consisting of patients with the lead implanted in the alternative position.Results: Of the 167 patients included in the study, 90 patients (54%) were assigned to the RVA group and 77 patients (46%) to the non-RVA group. The mean follow-up time was 26.1 12.6 months. Electrical parameters (pacing threshold, ventricular sensing amplitude and impedance) observed intraoperatively and measured on the first day following the procedure did not differ significantly between the groups. There was no significant difference in the incidence of abnormal electrical parameters between the apical and non-apical electrodes throughout the observation period. A total of 4 patients (4%) in the RVA group and 3 patients (4%) in the non-RVA group required lead replacement during the follow-up period.Conclusions: The position of the defibrillation lead has no clinically significant influence on the electrical parameters observed both intraoperatively and in long-term follow-up.</ns3:p>

Authors

Institutions

Publication Details

Journal
In a good rythm
Published
2024-09-27
DOI
https://doi.org/10.5604/01.3001.0054.7942
Primary Topic
Cardiac electrophysiology and arrhythmias
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The impact of defibrillation lead position on electrical parameters measured perioperatively and in long-term follow-up

Adam Wojtaszczyk, Bartosz Szepietowski, Michał Kałowski, Jerzy Krzysztof Wranicz et al.
In a good rythm
Cardiac electrophysiology and arrhythmias
article

The impact of defibrillation lead position on electrical parameters measured perioperatively and in long-term follow-up

Adam Wojtaszczyk, Bartosz Szepietowski, Michał Kałowski, Jerzy Krzysztof Wranicz, Paweł Ptaszyński, Krzysztof Kaczmarek
article en

Abstract

<ns3:p>Introduction: Implantable cardioverter-defibrillators (ICD) with transvenous defibrillation leads remain the leading therapeutic method in the treatment of patients at high risk of sudden cardiac death. The lack of clear data confirming the superiority of any position in the right ventricle for placement of the defibrillation lead results in significant individualization of practice between operators.Aims: The study aims to assess the electrical parameters of defibrillation leads obtained intraoperatively and in the long-term follow-up, depending on the position of the lead in the right ventricle.Methods: Between 2018 and 2021, consecutive patients were enrolled in an observational study after the implantation of a device with defibrillation capability. Electrical parameters were measured during each control visit, 3 months after the implantation procedure, and then every 6-months during follow-up visits. The data regarding intraprocedural course (procedure time, an X-ray time, intraprocedural electrical parameters) were collected retrospectively. The patients were divided into two groups depending on the placement of the defibrillation lead in the right ventricle: the RVA (right ventricular apex group) and the non-RVA group consisting of patients with the lead implanted in the alternative position.Results: Of the 167 patients included in the study, 90 patients (54%) were assigned to the RVA group and 77 patients (46%) to the non-RVA group. The mean follow-up time was 26.1 12.6 months. Electrical parameters (pacing threshold, ventricular sensing amplitude and impedance) observed intraoperatively and measured on the first day following the procedure did not differ significantly between the groups. There was no significant difference in the incidence of abnormal electrical parameters between the apical and non-apical electrodes throughout the observation period. A total of 4 patients (4%) in the RVA group and 3 patients (4%) in the non-RVA group required lead replacement during the follow-up period.Conclusions: The position of the defibrillation lead has no clinically significant influence on the electrical parameters observed both intraoperatively and in long-term follow-up.</ns3:p>

In a good rythmVol. 1(70)
Medical University of Lodz (PL)
Openalex Percentile: Top 67%
Cardiac electrophysiology and arrhythmias
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.