Investigation of mechanical performance for tricuspid valve stent in different compression modes

Abstract The tricuspid acting as a one-way valve between the right atrium and the right ventricle of the human heart, is very important for the blood circulation system. The regurgitation will be induced if the tricuspid cannot close normally, leading to symptoms such as arrhythmia and right heart failure. In the case of severe regurgitation, replacement with an artificial tricuspid valve instead of the natural tricuspid is the promising solution. For the typical artificial tricuspid valve, the bioprosthesis valve is fixed on a tricuspid valve stent which is usually compressed before being implanted into the heart. In this study, numerical simulation with an ABAQUS standard approach is carried out with a focus on the compression process of the tricuspid valve stent to evaluate the mechanical performances of two compression methods: one is the traditional radial compression method (RCM), and another is the axial compression method (ACM) that gradually compresses through a conical channel. The results indicate that the environment temperature is closely related to the stress distribution for both the RCM and ACM, in particular, a lower temperature can result in lower stress. The maximum strain is located at the inner corner of the joint of the stent ribs and the connection between the top arm and the stent, with a value over 12% at the temperature of 0°C, indicating the possible position for failure. The average diameter-stress curves of the RCM and ACM are respectively obtained for the compression process. To achieve long operation time, the RCM is recommended from the perspective of reducing compressive stress. The present study can provide the experience for the production and clinical application of tricuspid valve stents.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2024-10-01
DOI
https://doi.org/10.1088/1742-6596/2854/1/012112
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of mechanical performance for tricuspid valve stent in different compression modes

Mingkui Zhang, Xianwu Luo, Fengzhi Zhang, Zhonghua Xu et al.
Journal of Physics Conference Series
Cardiac Valve Diseases and Treatments
article

Investigation of mechanical performance for tricuspid valve stent in different compression modes

Mingkui Zhang, Xianwu Luo, Fengzhi Zhang, Zhonghua Xu, Wenhao Fang, Yongshun Zeng
article en

Abstract

Abstract The tricuspid acting as a one-way valve between the right atrium and the right ventricle of the human heart, is very important for the blood circulation system. The regurgitation will be induced if the tricuspid cannot close normally, leading to symptoms such as arrhythmia and right heart failure. In the case of severe regurgitation, replacement with an artificial tricuspid valve instead of the natural tricuspid is the promising solution. For the typical artificial tricuspid valve, the bioprosthesis valve is fixed on a tricuspid valve stent which is usually compressed before being implanted into the heart. In this study, numerical simulation with an ABAQUS standard approach is carried out with a focus on the compression process of the tricuspid valve stent to evaluate the mechanical performances of two compression methods: one is the traditional radial compression method (RCM), and another is the axial compression method (ACM) that gradually compresses through a conical channel. The results indicate that the environment temperature is closely related to the stress distribution for both the RCM and ACM, in particular, a lower temperature can result in lower stress. The maximum strain is located at the inner corner of the joint of the stent ribs and the connection between the top arm and the stent, with a value over 12% at the temperature of 0°C, indicating the possible position for failure. The average diameter-stress curves of the RCM and ACM are respectively obtained for the compression process. To achieve long operation time, the RCM is recommended from the perspective of reducing compressive stress. The present study can provide the experience for the production and clinical application of tricuspid valve stents.

Journal of Physics Conference SeriesVol. 2854(1)
Tsinghua University (CN)
Openalex Percentile: Top 67%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Investigation of mechanical performance for tricuspid valve stent in different compression modes — Mingkui Zhang, Xianwu Luo, et al. · Journal of Physics Conference Series (2024) | TGRS Research Map | TGRS