Exposure of Photostimulable Phosphor Plates with Red, Blue, and Green Light: Comparing Multiple Sequences Before Scanning

Background: This research aims to investigate the effect of changing the order of the colors by conducting 12 sequences of light exposure in red, green, or blue colors. Methods: Human-related material was not included. An Arduino-based experimental tool was developed with a 3D-printed light-tight platform. The photostimulable phosphor (PSP) plates were irradiated using the same parameters and metallic coins as contrast material. The plates were exposed to red, green, and blue colors, with variable color order and time intervals (20 seconds, 1 second, 100 milliseconds, 10 milliseconds). The total delay before scanning and total light exposure time was fixed, 12 sequences were conducted, and each experiment was repeated twice. Resulting radiographs were imported to ImageJ software and signal-to-noise (SNR) and contrast-tonoise (CNR) values were calculated by combining the ROIs dedicated for each contrast region. Results: SNR values were found in the range of 128.6-155.6, while CNR values were found between 63.2 and 74.7. The second experiment (R-B-G, 20 seconds) resulted in the lowest SNR (128.6) and CNR (63.2) values. In contrast, the twelfth experiment (R-B-G, 10 milliseconds) resulted in the highest SNR (155.6) and CNR (74.7) values. For 20 seconds, administration of the red color first resulted in greater signal loss than green and blue colors. Conclusion: Results of this experiment suggest that, despite the total delay and light exposure being the same, changing color order and time interval affected the resulting signal. Testing this phenomenon with different PSP systems or investigating the logic behind it may be the subject of future studies. Cite this article as: Amasya H, Orhan K. Exposure of photostimulable phosphor plates with red, blue, and green light: comparing multiple sequences before scanning. Essent Dent. 2024;3(3):94-99.

Authors

Publication Details

Journal
Essentials of Dentistry
Published
2024-10-23
DOI
https://doi.org/10.5152/essentdent.2024.24025
Primary Topic
Air Quality Monitoring and Forecasting
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Exposure of Photostimulable Phosphor Plates with Red, Blue, and Green Light: Comparing Multiple Sequences Before Scanning

Kaan Orhan, Hakan Amasya
Essentials of Dentistry
Air Quality Monitoring and Forecasting
article

Exposure of Photostimulable Phosphor Plates with Red, Blue, and Green Light: Comparing Multiple Sequences Before Scanning

Kaan Orhan, Hakan Amasya
article en

Abstract

Background: This research aims to investigate the effect of changing the order of the colors by conducting 12 sequences of light exposure in red, green, or blue colors. Methods: Human-related material was not included. An Arduino-based experimental tool was developed with a 3D-printed light-tight platform. The photostimulable phosphor (PSP) plates were irradiated using the same parameters and metallic coins as contrast material. The plates were exposed to red, green, and blue colors, with variable color order and time intervals (20 seconds, 1 second, 100 milliseconds, 10 milliseconds). The total delay before scanning and total light exposure time was fixed, 12 sequences were conducted, and each experiment was repeated twice. Resulting radiographs were imported to ImageJ software and signal-to-noise (SNR) and contrast-tonoise (CNR) values were calculated by combining the ROIs dedicated for each contrast region. Results: SNR values were found in the range of 128.6-155.6, while CNR values were found between 63.2 and 74.7. The second experiment (R-B-G, 20 seconds) resulted in the lowest SNR (128.6) and CNR (63.2) values. In contrast, the twelfth experiment (R-B-G, 10 milliseconds) resulted in the highest SNR (155.6) and CNR (74.7) values. For 20 seconds, administration of the red color first resulted in greater signal loss than green and blue colors. Conclusion: Results of this experiment suggest that, despite the total delay and light exposure being the same, changing color order and time interval affected the resulting signal. Testing this phenomenon with different PSP systems or investigating the logic behind it may be the subject of future studies. Cite this article as: Amasya H, Orhan K. Exposure of photostimulable phosphor plates with red, blue, and green light: comparing multiple sequences before scanning. Essent Dent. 2024;3(3):94-99.

Essentials of DentistryVol. 3(3)
Openalex Percentile: Top 83%
Air Quality Monitoring and Forecasting
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Exposure of Photostimulable Phosphor Plates with Red, Blue, and Green Light: Comparing Multiple Sequences Before Scanning — Kaan Orhan, Hakan Amasya · Essentials of Dentistry (2024) | TGRS Research Map | TGRS