Characterisation of an LED-based integrating sphere source for detection of changes of AERONET Europe radiometers

Abstract Within the framework of the EMPIR project 19ENV04 MAPP “Metrology for aerosol optical properties” an LED-based source was built for monitoring responsivity changes of Cimel filter radiometers from the AERONET Europe network. The developed LED device was flanged to integrating spheres to fill the field of view of the radiometers. Here we present the characterisation results of the LED-based integrating sphere source. The spectra of the radiant flux leaving a BaSO 4 -based integrating sphere showed positive wavelength dependent drift, which had not been reported in literature yet and did not appear by using incandescent lamps with the integrating spheres. This report shows approaches to reduce such drifts and an alternative using a PTFE-based integrating sphere, which was also tested for stability, reproducibility and homogeneity.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2024-10-01
DOI
https://doi.org/10.1088/1742-6596/2864/1/012017
Primary Topic
Calibration and Measurement Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterisation of an LED-based integrating sphere source for detection of changes of AERONET Europe radiometers

Peter Sperfeld, S. Pape, Saulius Nevas, K. M. Schwind
Journal of Physics Conference Series
Calibration and Measurement Techniques
article

Characterisation of an LED-based integrating sphere source for detection of changes of AERONET Europe radiometers

Peter Sperfeld, S. Pape, Saulius Nevas, K. M. Schwind
article en

Abstract

Abstract Within the framework of the EMPIR project 19ENV04 MAPP “Metrology for aerosol optical properties” an LED-based source was built for monitoring responsivity changes of Cimel filter radiometers from the AERONET Europe network. The developed LED device was flanged to integrating spheres to fill the field of view of the radiometers. Here we present the characterisation results of the LED-based integrating sphere source. The spectra of the radiant flux leaving a BaSO 4 -based integrating sphere showed positive wavelength dependent drift, which had not been reported in literature yet and did not appear by using incandescent lamps with the integrating spheres. This report shows approaches to reduce such drifts and an alternative using a PTFE-based integrating sphere, which was also tested for stability, reproducibility and homogeneity.

Journal of Physics Conference SeriesVol. 2864(1)
Physikalisch-Technische Bundesanstalt (DE)
Openalex Percentile: Top 71%
Calibration and Measurement Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.