Measurements of Low Optical Power with Cryostat-Based Predictable Quantum Efficient Detector at Liquid Nitrogen Temperature

Abstract We have validated optical power measurements with a Predictable Quantum Efficient Detector (PQED) at liquid nitrogen temperature (77 K) at low optical power from 130 fW to 3.3 pW. Two laser wavelengths at 514 nm and 785 nm were used. The lowest measured optical power corresponds to a photon flux of 0.5·10 6 photons per second (785 nm). The PQED’s responsivity is linear within the relative measurement uncertainties of 8% at 0.5·10 6 ph/s and 1.4% at 10·10 6 ph/s (95% confidence level), which enables the calibration of other low photon flux detectors directly against a primary standard of optical power.

Authors

Institutions

Publication Details

Journal
Journal of Physics Conference Series
Published
2024-10-01
DOI
https://doi.org/10.1088/1742-6596/2864/1/012005
Primary Topic
Advanced Optical Sensing Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Measurements of Low Optical Power with Cryostat-Based Predictable Quantum Efficient Detector at Liquid Nitrogen Temperature

Mikhail Korpusenko, Meelis Sildoja, Erkki Ikonen, Farshid Manoocheri
Journal of Physics Conference Series
Advanced Optical Sensing Technologies
article

Measurements of Low Optical Power with Cryostat-Based Predictable Quantum Efficient Detector at Liquid Nitrogen Temperature

Mikhail Korpusenko, Meelis Sildoja, Erkki Ikonen, Farshid Manoocheri
article en

Abstract

Abstract We have validated optical power measurements with a Predictable Quantum Efficient Detector (PQED) at liquid nitrogen temperature (77 K) at low optical power from 130 fW to 3.3 pW. Two laser wavelengths at 514 nm and 785 nm were used. The lowest measured optical power corresponds to a photon flux of 0.5·10 6 photons per second (785 nm). The PQED’s responsivity is linear within the relative measurement uncertainties of 8% at 0.5·10 6 ph/s and 1.4% at 10·10 6 ph/s (95% confidence level), which enables the calibration of other low photon flux detectors directly against a primary standard of optical power.

Journal of Physics Conference SeriesVol. 2864(1)
National metrology institute VTT MIKES (FI), Tallinn University (EE), National Institute of Chemical Physics and Biophysics (EE), Aalto University (FI)
Affordable and clean energy
Openalex Percentile: Top 85%
Advanced Optical Sensing Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.