Capillary drops, capillary pooled, and venous blood samples for determining hemoglobin concentration using HemoCue: A measurement system analysis

Several external and internal factors can affect the performance and variability of Hemoglobin concentration [Hb] measurements using HemoCue, and documentation on the contribution of different sources of [Hb] variation is limited. We used an experimental repeated measurements design with nine randomly selected participants, three HemoCue devices, and three trained field workers. HemoCue measurements for all samples were repeated three times. The [Hb] measurement system considers four sources of error: 1) HemoCue devices, 2) field workers, 3) between individuals, and 4) within individuals. A concordance analysis was used to assess accuracy and precision, and a linear mixed model was used to estimate mean differences (bias) from blood specimens, anticoagulants, and to estimate the contribution of the 4 sources of error to [Hb] measurements. Positive mean [Hb] differences were found: 1.34 g/dL for capillary drops, 0.81 g/dL for pooled capillary blood samples, 0.756 g/dL for venous blood stored with EDTA, and 0.911 g/dL for venous blood stored with heparin. The mean [Hb] difference for venous blood with EDTA was used as a correction factor for all results measured using a HemoCue. After adjustment, capillary drops showed a mean difference of 0.585 g/dL, and pooled capillary samples were not significantly different. The individual variability was 95.8% of total variance, HemoCue devices contributed 2.1% of measurement error, field staff contributed 0.4%, and the residual was 1.7%. The HemoCue [Hb] measurement system is reliable in controlled environments, with a small measurement error of 4.2%.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2024-10-25
DOI
https://doi.org/10.1371/journal.pone.0312233
Citations
3
Primary Topic
Iron Metabolism and Disorders
Type
article
Field-Weighted Citation Impact
1.36

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Capillary drops, capillary pooled, and venous blood samples for determining hemoglobin concentration using HemoCue: A measurement system analysis

Omar Dary, Teresa Shamah‐Levy, Vanessa De la Cruz‐Góngora, Ignacio Méndez‐Gómez‐Humarán
3 citations
PLoS ONE
Iron Metabolism and Disorders
1.36
article

Capillary drops, capillary pooled, and venous blood samples for determining hemoglobin concentration using HemoCue: A measurement system analysis

Omar Dary, Teresa Shamah‐Levy, Vanessa De la Cruz‐Góngora, Ignacio Méndez‐Gómez‐Humarán
article en
3 citations

Abstract

Several external and internal factors can affect the performance and variability of Hemoglobin concentration [Hb] measurements using HemoCue, and documentation on the contribution of different sources of [Hb] variation is limited. We used an experimental repeated measurements design with nine randomly selected participants, three HemoCue devices, and three trained field workers. HemoCue measurements for all samples were repeated three times. The [Hb] measurement system considers four sources of error: 1) HemoCue devices, 2) field workers, 3) between individuals, and 4) within individuals. A concordance analysis was used to assess accuracy and precision, and a linear mixed model was used to estimate mean differences (bias) from blood specimens, anticoagulants, and to estimate the contribution of the 4 sources of error to [Hb] measurements. Positive mean [Hb] differences were found: 1.34 g/dL for capillary drops, 0.81 g/dL for pooled capillary blood samples, 0.756 g/dL for venous blood stored with EDTA, and 0.911 g/dL for venous blood stored with heparin. The mean [Hb] difference for venous blood with EDTA was used as a correction factor for all results measured using a HemoCue. After adjustment, capillary drops showed a mean difference of 0.585 g/dL, and pooled capillary samples were not significantly different. The individual variability was 95.8% of total variance, HemoCue devices contributed 2.1% of measurement error, field staff contributed 0.4%, and the residual was 1.7%. The HemoCue [Hb] measurement system is reliable in controlled environments, with a small measurement error of 4.2%.

PLoS ONEVol. 19(10)
Mathematics Research Center (MX), Trinity College Dublin (IE), United States Agency for International Development (US), Instituto Nacional de Salud Pública (MX)
United States Agency for International Development, Secretaría de Salud
Openalex Percentile: Top 17%
Iron Metabolism and Disorders
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.