Integration of a dose management system in the public hospitals of Cyprus: Key areas and initial outcomes

Despite the growing use of dose management systems (DMSs), there is limited guidance on their optimal setup. This study aims to fill this gap by outlining the process followed to integrate "DOSE" (Qaelum NV, Belgium) DMS within the Cypriot public hospitals. DOSE was connected to the available central PACS, followed by a thorough data and functioning validation. A preliminary study was also conducted to assess the DOSE's ability to establish typical dose values in diagnostic radiography. Overall, over 70 x-ray units were configured. For some connections, it was necessary to first adjust the unit's DICOM attributes or to utilise the DICOM coercion technique. The DICOM image header was sometimes preferred over the DICOM RDSR, as the chosen data source. This study underscored the importance of understanding the equipment's technical capabilities for a successful DMS integration, while personnel with expertise in diagnostic radiology and the IT aspects of medical physics proved to have a vital role in the process. The pilot study showed that DOSE is an effective tool for establishing typical dose values and assessing compliance with relevant dose levels.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2024-10-24
DOI
https://doi.org/10.1093/rpd/ncae210
Citations
3
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
1.21
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Integration of a dose management system in the public hospitals of Cyprus: Key areas and initial outcomes

Vasileios I Metaxas, Georgiana Kokona, Demetris Kaolis, Eleni Skouridi
3 citations
Radiation Protection Dosimetry
Radiation Dose and Imaging
1.21
article

Integration of a dose management system in the public hospitals of Cyprus: Key areas and initial outcomes

Vasileios I Metaxas, Georgiana Kokona, Demetris Kaolis, Eleni Skouridi
article en
3 citations

Abstract

Despite the growing use of dose management systems (DMSs), there is limited guidance on their optimal setup. This study aims to fill this gap by outlining the process followed to integrate "DOSE" (Qaelum NV, Belgium) DMS within the Cypriot public hospitals. DOSE was connected to the available central PACS, followed by a thorough data and functioning validation. A preliminary study was also conducted to assess the DOSE's ability to establish typical dose values in diagnostic radiography. Overall, over 70 x-ray units were configured. For some connections, it was necessary to first adjust the unit's DICOM attributes or to utilise the DICOM coercion technique. The DICOM image header was sometimes preferred over the DICOM RDSR, as the chosen data source. This study underscored the importance of understanding the equipment's technical capabilities for a successful DMS integration, while personnel with expertise in diagnostic radiology and the IT aspects of medical physics proved to have a vital role in the process. The pilot study showed that DOSE is an effective tool for establishing typical dose values and assessing compliance with relevant dose levels.

Radiation Protection DosimetryVol. 200(20)
Nicosia General Hospital (CY)
Openalex Percentile: Top 19%
Radiation Dose and Imaging
1.21
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Integration of a dose management system in the public hospitals of Cyprus: Key areas and initial outcomes — Vasileios I Metaxas, Georgiana Kokona, et al. · Radiation Protection Dosimetry (2024) | TGRS Research Map | TGRS