Validation of the TMB-2296-BT blood pressure monitor in adults according to the ISO 81060-2:2018 + Amd.1:2020

This study aimed to validate the accuracy of the test device (TMB-2296-BT) blood pressure (BP) monitor in adults according to International Standard Organization (ISO) 81060-2:2018 + Amd.1:2020 universal standard protocol, which is a digital monitor. Three trained observers used the same arm sequential method to compare the SBPs and DBPs measured by the test device with those measured by the reference device (mercury sphygmomanometer). For the test device with cuff ranging from 22 to 32 cm, there are 88 adults, with a male-to-female ratio of 35 : 53. The mean difference and SD between reference BPs and test device BPs readings were (0.21 ± 2.59)/(0.66 ± 2.12) mmHg for SBP/DBP for criterion 1, and (0.21 ± 2.07)/(0.66 ± 1.76) mmHg for SBP/DBP for criterion 2. For the test device with cuffs ranging from 22 to 42 cm, there are 87 adults, with a male-to-female ratio of 49 : 38. The mean difference and SD between reference BPs and test device BPs readings were (-1.62 ± 2.80)/(0.12 ± 3.01) mmHg for SBP/DBP for criterion 1, and (-1.62 ± 2.35)/(0.12 ± 2.60) mmHg for SBP/ DBP for criterion 2. And for the test device with cuff arm circumference sizes of 22-32 cm and 22-42 cm fulfilled both validation criterion 1 and 2 of the ISO 81060-2:2018 + Amd.1:2020 standard and can be recommended for both clinical and self/home BP measurement in adults.

Authors

Institutions

Publication Details

Journal
Blood Pressure Monitoring
Published
2024-10-22
DOI
https://doi.org/10.1097/mbp.0000000000000728
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the TMB-2296-BT blood pressure monitor in adults according to the ISO 81060-2:2018 + Amd.1:2020

Chaoya Li, Zhu Li, Bin Peng, Jiahui Liang et al.
Blood Pressure Monitoring
Blood Pressure and Hypertension Studies
article

Validation of the TMB-2296-BT blood pressure monitor in adults according to the ISO 81060-2:2018 + Amd.1:2020

Chaoya Li, Zhu Li, Bin Peng, Jiahui Liang, Jia Hu, Xiao Zheng, Xinda Wang, Zijian Xie, Hua Deng, Xiaoqin Du
article en

Abstract

This study aimed to validate the accuracy of the test device (TMB-2296-BT) blood pressure (BP) monitor in adults according to International Standard Organization (ISO) 81060-2:2018 + Amd.1:2020 universal standard protocol, which is a digital monitor. Three trained observers used the same arm sequential method to compare the SBPs and DBPs measured by the test device with those measured by the reference device (mercury sphygmomanometer). For the test device with cuff ranging from 22 to 32 cm, there are 88 adults, with a male-to-female ratio of 35 : 53. The mean difference and SD between reference BPs and test device BPs readings were (0.21 ± 2.59)/(0.66 ± 2.12) mmHg for SBP/DBP for criterion 1, and (0.21 ± 2.07)/(0.66 ± 1.76) mmHg for SBP/DBP for criterion 2. For the test device with cuffs ranging from 22 to 42 cm, there are 87 adults, with a male-to-female ratio of 49 : 38. The mean difference and SD between reference BPs and test device BPs readings were (-1.62 ± 2.80)/(0.12 ± 3.01) mmHg for SBP/DBP for criterion 1, and (-1.62 ± 2.35)/(0.12 ± 2.60) mmHg for SBP/ DBP for criterion 2. And for the test device with cuff arm circumference sizes of 22-32 cm and 22-42 cm fulfilled both validation criterion 1 and 2 of the ISO 81060-2:2018 + Amd.1:2020 standard and can be recommended for both clinical and self/home BP measurement in adults.

Blood Pressure MonitoringVol. 30(3)
Chenzhou First People's Hospital (CN)
Quality Education
Openalex Percentile: Top 68%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Validation of the TMB-2296-BT blood pressure monitor in adults according to the ISO 81060-2:2018 + Amd.1:2020 — Chaoya Li, Zhu Li, et al. · Blood Pressure Monitoring (2024) | TGRS Research Map | TGRS