Enhancing Total Temperature Measurement Accuracy: Calibration Procedures and Novel Two-Wire Probes

Abstract Total temperature measurement with thermocouple probes suffers from errors due to heat transfer effects. Two dominant sources of errors are convection and conduction between the thermocouple, the support, and the flow. These effects can be treated in two different categories: the velocity error, created by convection from the internal flow velocity in the probe shield, and the conduction error, involving heat transfer through the wire to the shield and the probe stem due to temperature differences between each component. This article presents a robust approach to experimentally assess and reduce both errors. An open jet test stand at the Purdue Experimental Turbine Aerothermal Laboratory is used to evaluate the effects of the velocity error at various Mach numbers. Infrared (IR) thermography measurements have been conducted to assess temperature gradients on the probe support during this calibration. With this information, it is possible to correct for conduction errors during the calibration and obtain a recovery factor that is solely dependent on upstream velocity. Recovery factor of 0.94±0.05 for Mach 0.2 and 0.98±0.005 for Mach 0.9 is obtained. Progress has been made on a two-wire thermocouple approach to address conduction errors in the testing scenario, where IR thermography is not possible. The readings from two wires with different length-to-diameter ratios are used with a novel linear correction method to correct for the flow total temperature. Advances in the design and manufacturing of the probe are presented, facilitating faster design iterations and implementation, and providing full control over the calibration procedures. This method can correct conduction errors within 0.2 K for Mach 0.6.

Authors

Institutions

Publication Details

Journal
Journal of Turbomachinery
Published
2024-10-22
DOI
https://doi.org/10.1115/1.4066964
Primary Topic
Advanced Sensor Technologies Research
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing Total Temperature Measurement Accuracy: Calibration Procedures and Novel Two-Wire Probes

Guillermo Paniagua, Diego Sánchez de la Rosa
Journal of Turbomachinery
Advanced Sensor Technologies Research
article

Enhancing Total Temperature Measurement Accuracy: Calibration Procedures and Novel Two-Wire Probes

Guillermo Paniagua, Diego Sánchez de la Rosa
article en

Abstract

Abstract Total temperature measurement with thermocouple probes suffers from errors due to heat transfer effects. Two dominant sources of errors are convection and conduction between the thermocouple, the support, and the flow. These effects can be treated in two different categories: the velocity error, created by convection from the internal flow velocity in the probe shield, and the conduction error, involving heat transfer through the wire to the shield and the probe stem due to temperature differences between each component. This article presents a robust approach to experimentally assess and reduce both errors. An open jet test stand at the Purdue Experimental Turbine Aerothermal Laboratory is used to evaluate the effects of the velocity error at various Mach numbers. Infrared (IR) thermography measurements have been conducted to assess temperature gradients on the probe support during this calibration. With this information, it is possible to correct for conduction errors during the calibration and obtain a recovery factor that is solely dependent on upstream velocity. Recovery factor of 0.94±0.05 for Mach 0.2 and 0.98±0.005 for Mach 0.9 is obtained. Progress has been made on a two-wire thermocouple approach to address conduction errors in the testing scenario, where IR thermography is not possible. The readings from two wires with different length-to-diameter ratios are used with a novel linear correction method to correct for the flow total temperature. Advances in the design and manufacturing of the probe are presented, facilitating faster design iterations and implementation, and providing full control over the calibration procedures. This method can correct conduction errors within 0.2 K for Mach 0.6.

Journal of TurbomachineryVol. 147(5)
Purdue University West Lafayette (US)
Openalex Percentile: Top 83%
Advanced Sensor Technologies Research
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.