Removal of errant coils using the Tigertriever-13
Coil migration poses a significant intraprocedural risk during coil embolization in interventional neuroradiology procedures. In this technical video we describe the technique of removal of errand coils using the low profile Tigertriever-13 device.1 Traditional methods of errant coil retrieval rely on the use of stent retrievers;2 however, these devices are usually too large to fit through smaller inner diameter 0.013-0.0165 inch coiling catheters, thus requiring upsizing materials in emergency situations.3 We demonstrate the technique of coil retrieval using this low-profile device ex-vivo in silicone flow models, then its application in vivo as a 'bail-out' technique during acute embolization of a ruptured peri-callosal aneurysm which required retrieval of errant coils (video 1). neurintsurg;16/12/1212/V1F1V1Video 1Errant coil retrieval using the Tigertriever device.
Authors
- Emily Chung (ORCID: https://orcid.org/0000-0002-1326-8693)
- Pascal J. Mosimann (ORCID: https://orcid.org/0000-0001-5111-1972)
- Hugo Boechat Andrade
- Andrew Falzon (ORCID: https://orcid.org/0000-0003-2212-6635)
- Shah Islam (ORCID: https://orcid.org/0000-0002-9013-3717)
- Ze'ev Itzakzonhayosh
Institutions
- University Health Network (CA)
- Toronto Western Hospital (CA)
Publication Details
- Journal
- Journal of NeuroInterventional Surgery
- Published
- 2024-10-20
- DOI
- https://doi.org/10.1136/jnis-2024-021923
- Primary Topic
- Intracranial Aneurysms: Treatment and Complications
- Type
- article
- Field-Weighted Citation Impact