A Method for Compensating Hemoglobin Interference in Total Serum Bilirubin Measurement Using a Simple Two-Wavelength Reflectance Photometer

Neonatal hyperbilirubinemia (NH) is a common condition in newborns, with elevated bilirubin levels potentially causing neurological damage or death. Accurate and timely measurements of total serum bilirubin are essential to prevent these outcomes. Direct spectrophotometry, a reliable method for measuring bilirubin, is particularly useful in constrained settings due to its potential for portable low-cost instrumentation. However, this method is susceptible to interference from hemoglobin, often present due to hemolysis. Typically, this interference is reduced using complex optical filters, reagents, multiple wavelengths, or combinations thereof, which increase costs and complexity while reducing usability. This study presents a hemoglobin compensation algorithm applied to a simple, portable, two-wavelength (465 and 590 nm) reflectance photometer designed to receive 30 µL of plasma or whole blood samples and perform the measurement without any reagents. Testing across five bilirubin and hemoglobin levels (4.96 to 28 mg/dL and 0.06 to 0.99 g/dL, respectively) demonstrated that the algorithm effectively reduces hemoglobin interference and overestimation errors. The overall root mean square error was reduced from 4.86 to 1.45 mg/dL, while the measurement bias decreased from -4.46 to -0.10 mg/dL. This substantial reduction in overestimation errors supports future clinical trials with neonatal blood samples.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2024-10-20
DOI
https://doi.org/10.3390/s24206749
Citations
1
Primary Topic
Neonatal Health and Biochemistry
Type
article
Field-Weighted Citation Impact
1.11
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Method for Compensating Hemoglobin Interference in Total Serum Bilirubin Measurement Using a Simple Two-Wavelength Reflectance Photometer

Carlos D. Coda Zabetta, Agostino Accardo, Miloš Ajčević, Lorenzo Zucchini
1 citations
Sensors
Neonatal Health and Biochemistry
1.11
article

A Method for Compensating Hemoglobin Interference in Total Serum Bilirubin Measurement Using a Simple Two-Wavelength Reflectance Photometer

Carlos D. Coda Zabetta, Agostino Accardo, Miloš Ajčević, Lorenzo Zucchini
article en
1 citations

Abstract

Neonatal hyperbilirubinemia (NH) is a common condition in newborns, with elevated bilirubin levels potentially causing neurological damage or death. Accurate and timely measurements of total serum bilirubin are essential to prevent these outcomes. Direct spectrophotometry, a reliable method for measuring bilirubin, is particularly useful in constrained settings due to its potential for portable low-cost instrumentation. However, this method is susceptible to interference from hemoglobin, often present due to hemolysis. Typically, this interference is reduced using complex optical filters, reagents, multiple wavelengths, or combinations thereof, which increase costs and complexity while reducing usability. This study presents a hemoglobin compensation algorithm applied to a simple, portable, two-wavelength (465 and 590 nm) reflectance photometer designed to receive 30 µL of plasma or whole blood samples and perform the measurement without any reagents. Testing across five bilirubin and hemoglobin levels (4.96 to 28 mg/dL and 0.06 to 0.99 g/dL, respectively) demonstrated that the algorithm effectively reduces hemoglobin interference and overestimation errors. The overall root mean square error was reduced from 4.86 to 1.45 mg/dL, while the measurement bias decreased from -4.46 to -0.10 mg/dL. This substantial reduction in overestimation errors supports future clinical trials with neonatal blood samples.

SensorsVol. 24(20)
University of Trieste (IT), Elettra-Sincrotrone Trieste S.C.p.A. (IT)
Good health and well-being
Openalex Percentile: Top 17%
Neonatal Health and Biochemistry
1.11
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Method for Compensating Hemoglobin Interference in Total Serum Bilirubin Measurement Using a Simple Two-Wavelength Reflectance Photometer — Carlos D. Coda Zabetta, Agostino Accardo, et al. · Sensors (2024) | TGRS Research Map | TGRS