Detection of carbapenem-resistant gram-negative bacilli in Japan using the fully automated bacterial testing device RAISUS S4
When F-GNR were based on MEPM MIC ≥ 0.25 µg/mL, CP-GNR were detected with a sensitivity of 100% (58/58) for the 18-h method and 98.3% (57/58) for the rapid method; the shortest detection times were 5.3 and 4.0 h, respectively. When NF-GNR were based on MEPM MIC > 8 µg/mL, it was possible to detect CP-GNR with 100% sensitivity (58/58) in both methods. Furthermore, in the analysis using the 18-h method for monitoring resistant bacteria, when ≥ 2 µg/mL was used as the screening concentration for F-GNR, approximately 50% of the resistant genotypes, NDM, GES, and KPC, were detected in approximately 7 h However, detecting the IMP and VIM took 11-12 h CONCLUSIONS: The 18-h and rapid methods with RAISUS S4 were highly correlated with the results of the microdilution method of CLSI, and CP-GNR detection was rapid using a function that can monitor the growth curve with RAISUS S4.
Authors
- Kaoru Tominaga (ORCID: https://orcid.org/0000-0001-8799-528X)
- Yumiko Funashima
- Takahiro Yaguchi (ORCID: https://orcid.org/0000-0002-4660-5123)
- Tsukuru Umemura (ORCID: https://orcid.org/0000-0003-1187-3248)
- Zenzo Nagasawa
- Kentaro Wakamatsu (ORCID: https://orcid.org/0000-0003-4311-887X)
- Rin Hamabe
Institutions
- Omuta hospital (JP)
Publication Details
- Journal
- Journal of Global Antimicrobial Resistance
- Published
- 2024-10-18
- DOI
- https://doi.org/10.1016/j.jgar.2024.09.009
- Citations
- 1
- Primary Topic
- Antibiotic Resistance in Bacteria
- Type
- article
- Field-Weighted Citation Impact
- 0.17