Detection of carbapenem-resistant gram-negative bacilli in Japan using the fully automated bacterial testing device RAISUS S4

When F-GNR were based on MEPM MIC ≥ 0.25 µg/mL, CP-GNR were detected with a sensitivity of 100% (58/58) for the 18-h method and 98.3% (57/58) for the rapid method; the shortest detection times were 5.3 and 4.0 h, respectively. When NF-GNR were based on MEPM MIC > 8 µg/mL, it was possible to detect CP-GNR with 100% sensitivity (58/58) in both methods. Furthermore, in the analysis using the 18-h method for monitoring resistant bacteria, when ≥ 2 µg/mL was used as the screening concentration for F-GNR, approximately 50% of the resistant genotypes, NDM, GES, and KPC, were detected in approximately 7 h However, detecting the IMP and VIM took 11-12 h CONCLUSIONS: The 18-h and rapid methods with RAISUS S4 were highly correlated with the results of the microdilution method of CLSI, and CP-GNR detection was rapid using a function that can monitor the growth curve with RAISUS S4.

Authors

Institutions

Publication Details

Journal
Journal of Global Antimicrobial Resistance
Published
2024-10-18
DOI
https://doi.org/10.1016/j.jgar.2024.09.009
Citations
1
Primary Topic
Antibiotic Resistance in Bacteria
Type
article
Field-Weighted Citation Impact
0.17
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Detection of carbapenem-resistant gram-negative bacilli in Japan using the fully automated bacterial testing device RAISUS S4

Kaoru Tominaga, Yumiko Funashima, Takahiro Yaguchi, Tsukuru Umemura et al.
1 citations
Journal of Global Antimicrobial Resistance
Antibiotic Resistance in Bacteria
0.17
article

Detection of carbapenem-resistant gram-negative bacilli in Japan using the fully automated bacterial testing device RAISUS S4

Kaoru Tominaga, Yumiko Funashima, Takahiro Yaguchi, Tsukuru Umemura, Zenzo Nagasawa, Kentaro Wakamatsu, Rin Hamabe
article en
1 citations

Abstract

When F-GNR were based on MEPM MIC ≥ 0.25 µg/mL, CP-GNR were detected with a sensitivity of 100% (58/58) for the 18-h method and 98.3% (57/58) for the rapid method; the shortest detection times were 5.3 and 4.0 h, respectively. When NF-GNR were based on MEPM MIC > 8 µg/mL, it was possible to detect CP-GNR with 100% sensitivity (58/58) in both methods. Furthermore, in the analysis using the 18-h method for monitoring resistant bacteria, when ≥ 2 µg/mL was used as the screening concentration for F-GNR, approximately 50% of the resistant genotypes, NDM, GES, and KPC, were detected in approximately 7 h However, detecting the IMP and VIM took 11-12 h CONCLUSIONS: The 18-h and rapid methods with RAISUS S4 were highly correlated with the results of the microdilution method of CLSI, and CP-GNR detection was rapid using a function that can monitor the growth curve with RAISUS S4.

Journal of Global Antimicrobial ResistanceVol. 39
Omuta hospital (JP)
Openalex Percentile: Top 56%
Antibiotic Resistance in Bacteria
0.17
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Detection of carbapenem-resistant gram-negative bacilli in Japan using the fully automated bacterial testing device RAISUS S4 — Kaoru Tominaga, Yumiko Funashima, et al. · Journal of Global Antimicrobial Resistance (2024) | TGRS Research Map | TGRS