The use of a novel toe–thumb pressure index for assessing arterial status in the lower limb. A reliability and validity study
Thumb systolic pressures are correlated with brachial systolic pressures, with reasonable intra-rater reliability, however, correlation is only fair and measurement error wider than clinically acceptable. Furthermore, manual measures are poorly correlated with automated units. Consequently, caution is required in applying these techniques. As thumb measures were perceived as significantly more comfortable than brachial measures and have an advantage where brachial pressures cannot, or should not, be obtained, further evaluation is warranted.