Comprehensive analysis of peripheral dose in electron beam therapy with a Varian TrueBeam Linac

To investigate the characteristics of peripheral doses outside electron-beam applicators in Varian TrueBeam linacs. Peripheral doses outside the electron applicator were measured for 6-, 9- and 12-MeV beams at the maximum dose depth ( D max ) for each energy source and at a source-to-surface distance (SSD) of 100 cm. Measurements were performed using EBT3 films in solid water phantoms. The impact of field size on the penumbra width and peripheral doses was studied using various cutouts, including 3 ​cm ​× ​3 ​cm, 6 ​cm ​× ​6 ​cm, and 10 ​cm ​× ​10 ​cm in a 10 ​cm ​× ​10 ​cm applicator with the gantry and collimator at 0°. The influence of the applicator size was investigated using a circular cutout of 5 ​cm in diameter for various applicator sizes, including 6 ​cm ​× ​6 ​cm, 10 ​cm ​× ​10 ​cm, 15 ​cm ​× ​15 ​cm, 20 cm ​× ​20 ​cm, and 25 ​cm ​× ​25 ​cm, at D max for each energy, while keeping the gantry and collimator angle at 0°. The measured dose profiles were compared with the Eclipse treatment planning system (TPS) predicted dose profiles. The effect of varying gantry angles (0°, 90°, and 270°) for a 3 ​cm ​× ​3 ​cm cutout in a 10 ​cm ​× ​10 ​cm applicator for each energy source and varying collimator angles (0°, 90°, and 270°) for a 10 ​cm ​× ​10 ​cm field were investigated to determine their effects on the penumbra widths and peripheral doses. Both the penumbra width and peripheral dose values increased with energy across different field sizes, gantry angles, collimator angles, and applicator sizes. Root mean square deviation (RMSD) analysis indicated minimal differences between the measured profiles and TPS data. Peripheral doses remained below 5% of the maximum dose approximately 10–15 ​mm away from the field edges, suggesting the potential for implementing additional shielding where required. This study highlights the importance of considering peripheral doses in electron radiotherapy. It is important to note the impact on healthy tissues beyond the treatment area to ensure patient safety and prevent the long-term side effects of treatment. These findings emphasize the necessity of implementing appropriate measures to minimize peripheral doses.

Authors

Publication Details

Journal
Radiation Medicine and Protection
Published
2024-10-18
DOI
https://doi.org/10.1016/j.radmp.2024.10.003
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comprehensive analysis of peripheral dose in electron beam therapy with a Varian TrueBeam Linac

Joshua Dass, Pejman Rowshanfarzad, Mahsheed Sabet, Kapil Dev Maharaj et al.
Radiation Medicine and Protection
Advanced Radiotherapy Techniques
article

Comprehensive analysis of peripheral dose in electron beam therapy with a Varian TrueBeam Linac

Joshua Dass, Pejman Rowshanfarzad, Mahsheed Sabet, Kapil Dev Maharaj, Mounir Ibrahim, Talat Mahmood
article en

Abstract

To investigate the characteristics of peripheral doses outside electron-beam applicators in Varian TrueBeam linacs. Peripheral doses outside the electron applicator were measured for 6-, 9- and 12-MeV beams at the maximum dose depth ( D max ) for each energy source and at a source-to-surface distance (SSD) of 100 cm. Measurements were performed using EBT3 films in solid water phantoms. The impact of field size on the penumbra width and peripheral doses was studied using various cutouts, including 3 ​cm ​× ​3 ​cm, 6 ​cm ​× ​6 ​cm, and 10 ​cm ​× ​10 ​cm in a 10 ​cm ​× ​10 ​cm applicator with the gantry and collimator at 0°. The influence of the applicator size was investigated using a circular cutout of 5 ​cm in diameter for various applicator sizes, including 6 ​cm ​× ​6 ​cm, 10 ​cm ​× ​10 ​cm, 15 ​cm ​× ​15 ​cm, 20 cm ​× ​20 ​cm, and 25 ​cm ​× ​25 ​cm, at D max for each energy, while keeping the gantry and collimator angle at 0°. The measured dose profiles were compared with the Eclipse treatment planning system (TPS) predicted dose profiles. The effect of varying gantry angles (0°, 90°, and 270°) for a 3 ​cm ​× ​3 ​cm cutout in a 10 ​cm ​× ​10 ​cm applicator for each energy source and varying collimator angles (0°, 90°, and 270°) for a 10 ​cm ​× ​10 ​cm field were investigated to determine their effects on the penumbra widths and peripheral doses. Both the penumbra width and peripheral dose values increased with energy across different field sizes, gantry angles, collimator angles, and applicator sizes. Root mean square deviation (RMSD) analysis indicated minimal differences between the measured profiles and TPS data. Peripheral doses remained below 5% of the maximum dose approximately 10–15 ​mm away from the field edges, suggesting the potential for implementing additional shielding where required. This study highlights the importance of considering peripheral doses in electron radiotherapy. It is important to note the impact on healthy tissues beyond the treatment area to ensure patient safety and prevent the long-term side effects of treatment. These findings emphasize the necessity of implementing appropriate measures to minimize peripheral doses.

Radiation Medicine and ProtectionVol. 6(1)
Openalex Percentile: Top 74%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.