Design and Use of a Custom Phantom for Regular Tests of Radiography Apparatus: A Feasibility Study

This study investigates the feasibility of employing an in-house-developed physical phantom dedicated to the weekly quality control testing of radiographic systems, performed by radiographers. For this purpose, a 3D phantom was fabricated, featuring test objects, including a model representing a lesion. Alongside this phantom, a commercial phantom, specifically, IBA's Primus L, was utilized. Weekly imaging of both phantoms was conducted over a span of four weeks, involving different imaging protocols and anode voltages. Subsequently, the obtained data underwent visual evaluation, as well as measurement of the intensity of selected regions of interest. The average values for three incident kilovoltages remained consistently stable over the four weeks, with the exception of the "low energy" case, which exhibited variability during the first week of measurements. Following experiments in "Week 1", the X-Ray unit was identified as malfunctioning and underwent necessary repairs. The in-house-developed phantom demonstrated its utility in assessing the performance of the X-Ray system.

Authors

Institutions

Publication Details

Journal
Journal of Imaging
Published
2024-10-18
DOI
https://doi.org/10.3390/jimaging10100258
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Design and Use of a Custom Phantom for Regular Tests of Radiography Apparatus: A Feasibility Study

Kristina Bliznakova, Virginia Tsapaki, Nikolay Dukov, Mariana Yordanova et al.
Journal of Imaging
Radiation Dose and Imaging
article

Design and Use of a Custom Phantom for Regular Tests of Radiography Apparatus: A Feasibility Study

Kristina Bliznakova, Virginia Tsapaki, Nikolay Dukov, Mariana Yordanova, Vanessa-Mery Valkova
article en

Abstract

This study investigates the feasibility of employing an in-house-developed physical phantom dedicated to the weekly quality control testing of radiographic systems, performed by radiographers. For this purpose, a 3D phantom was fabricated, featuring test objects, including a model representing a lesion. Alongside this phantom, a commercial phantom, specifically, IBA's Primus L, was utilized. Weekly imaging of both phantoms was conducted over a span of four weeks, involving different imaging protocols and anode voltages. Subsequently, the obtained data underwent visual evaluation, as well as measurement of the intensity of selected regions of interest. The average values for three incident kilovoltages remained consistently stable over the four weeks, with the exception of the "low energy" case, which exhibited variability during the first week of measurements. Following experiments in "Week 1", the X-Ray unit was identified as malfunctioning and underwent necessary repairs. The in-house-developed phantom demonstrated its utility in assessing the performance of the X-Ray system.

Journal of ImagingVol. 10(10)
International Atomic Energy Agency (AT), Medical University of Varna (BG)
Affordable and clean energy
Openalex Percentile: Top 70%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.