Characterization of brass mesh bolus for electron beam therapy

<i>Purpose</i>. Bolus is often required for targets close to or on skin surface, however, standard bolus on complex surfaces can result in air gaps that compromise dosimetry. Brass mesh boluses (RPD, Inc., Albertville, MN) are designed to conform to the patient's surface and reduce air gaps. While they have been well characterized for their use with photons, minimal characterization exists in literature for their use with electrons.<i>Methods and materials.</i>Dosimetric characteristics of brass mesh bolus was investigated for use with 6, 9 and 12 MeV electrons using a 10 × 10 cm<sup>2</sup>applicator on standard multi-energy LINAC. Measurements for bolus equivalence and percentage depth doses (PDDs) under brass mesh, as well as surface dose measurements were performed on solid water and a 3D printed resin breast phantom (Anycubic Photon MonoX, Shenzhen, China) using Markus<sup>®</sup>parallel-plate ionization chamber (Model 34045, PTW Freiburg, Germany), thermoluminescent detectors (TLD) and EBRT film. After obtaining surface dose measurements, these were compared to dose calculated on the Pinnacle3 treatment planning system (TPS, 16.2, Koninklijke Philips N.V.).<i>Results</i>. Measurements of surface dose under brass mesh showed consistently higher dose than without bolus, confirming that brass mesh can increase the PDD at surface up to ∼ 94% of dose at d<sub>max</sub>, depending on incident electron energy. This increase is equivalent to using ∼ 7.2 mm water equivalent bolus for 6 MeV, ∼ 3.6 mm for 9 MeV and ∼ 2.2 mm bolus for 12 MeV electrons. TPS results showed close agreement with<i>in-vivo</i>measurements, confirming the potential for brass mesh as bolus for electron irradiation, provided blousing effect is correctly modelled.<i>Conclusions</i>. To increase electron surface dose, a brass mesh can be used with equivalent effect of water-density bolus varying with electron energy. Proper implementation could allow for ease of treatment, as well as increase bolus conformality in electron-only plans.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2024-10-17
DOI
https://doi.org/10.1088/2057-1976/ad87f7
Citations
1
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.57
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Characterization of brass mesh bolus for electron beam therapy

Sara Lim, Slade Klawikowski, Eric D. Donnelly, Indra J. Das et al.
1 citations
Biomedical Physics & Engineering Express
Orthopaedic implants and arthroplasty
0.57
article

Characterization of brass mesh bolus for electron beam therapy

Sara Lim, Slade Klawikowski, Eric D. Donnelly, Indra J. Das, John P. Hayes, James J. Sohn
article en
1 citations

Abstract

<i>Purpose</i>. Bolus is often required for targets close to or on skin surface, however, standard bolus on complex surfaces can result in air gaps that compromise dosimetry. Brass mesh boluses (RPD, Inc., Albertville, MN) are designed to conform to the patient's surface and reduce air gaps. While they have been well characterized for their use with photons, minimal characterization exists in literature for their use with electrons.<i>Methods and materials.</i>Dosimetric characteristics of brass mesh bolus was investigated for use with 6, 9 and 12 MeV electrons using a 10 × 10 cm<sup>2</sup>applicator on standard multi-energy LINAC. Measurements for bolus equivalence and percentage depth doses (PDDs) under brass mesh, as well as surface dose measurements were performed on solid water and a 3D printed resin breast phantom (Anycubic Photon MonoX, Shenzhen, China) using Markus<sup>®</sup>parallel-plate ionization chamber (Model 34045, PTW Freiburg, Germany), thermoluminescent detectors (TLD) and EBRT film. After obtaining surface dose measurements, these were compared to dose calculated on the Pinnacle3 treatment planning system (TPS, 16.2, Koninklijke Philips N.V.).<i>Results</i>. Measurements of surface dose under brass mesh showed consistently higher dose than without bolus, confirming that brass mesh can increase the PDD at surface up to ∼ 94% of dose at d<sub>max</sub>, depending on incident electron energy. This increase is equivalent to using ∼ 7.2 mm water equivalent bolus for 6 MeV, ∼ 3.6 mm for 9 MeV and ∼ 2.2 mm bolus for 12 MeV electrons. TPS results showed close agreement with<i>in-vivo</i>measurements, confirming the potential for brass mesh as bolus for electron irradiation, provided blousing effect is correctly modelled.<i>Conclusions</i>. To increase electron surface dose, a brass mesh can be used with equivalent effect of water-density bolus varying with electron energy. Proper implementation could allow for ease of treatment, as well as increase bolus conformality in electron-only plans.

Biomedical Physics & Engineering ExpressVol. 10(6)
Northwestern University (US), City Of Hope National Medical Center (US), Medical College of Wisconsin (US), University of Chicago (US), City of Hope (US), Northwestern Medicine (US)
Openalex Percentile: Top 29%
Orthopaedic implants and arthroplasty
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.