Signal-to-noise trade-offs between magnet diameter and shield-to-coil distance for cylindrical Halbach-based portable MRI systems for neuroimaging
This work presents an overview of the trade-offs in transmit efficiencies for RF coils used for POC MRI neuroimaging as a function of coil-to-shield distance and inner diameter of the Halbach magnet. Results show that there is a relatively shallow optimum between a magnet diameter of 290 and 330 mm, with values falling more than 10% if either smaller or larger magnets are used.