A Study on Optimizing the Smoothing Factor for Fast Non-local Means-based Denoising Algorithm in Brain CT Phantom Image using 3D Printing

The purpose of this study is to fabricate a 3D printed phantom for computed tomography (CT) imaging and to optimize the smoothing factor of the fast non-local means (FNLM) denoising algorithm. For this purpose, a 3D printer with fused filament fabrication technology was used to produce the brain CT phantom with a diameter of 16 cm that mimics cerebrospinal fluid, gray matter, white matter, and bone. Then, we added the Gaussian noise of 0.005 to express the low-dose conditions from CT images acquired using a brain phantom. The optimization of performed by applying the FNLM algorithm to the noise-added CT phantom image. The d-value for adjusting smoothing intensity of FNLM algorithm was applied from 0.01 to 1.00 in increments of 0.01. The contrast to noise ratio (CNR), signal to noise ratio (SNR), and pixel profiles were measured for quantitative evaluation. As a result, both CNR and SNR were greatly improved until the d-value was 0.07, however, gradual changes were observed showing a flattened slope after 0.07. In addition, we confirmed that the signal value was stabilized at 0.07 in the measurement of the pixel profile. The incline exhibited a milder at subsequent d-values and the contrast was decreased. Based on these results, the smoothing factor of the FNLM algorithm was optimized to 0.07. Additionally, a comparative evaluation with the Wiener filter, median modified Wiener filter and total variation algorithm was conducted. As a result, the most improved CNR and SNR values were obtained in all tissues when applying the FNLM algorithm with an optimized d-value of 0.07. In conclusion, the FNLM algorithm with d-value of 0.07 was verified to be effective in reducing noise for low-dose CT conditions.

Authors

Publication Details

Journal
Daehan CT yeongsang gisul hakoeji
Published
2024-09-30
DOI
https://doi.org/10.31320/jksct.2024.26.2.43
Primary Topic
Medical Imaging and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Study on Optimizing the Smoothing Factor for Fast Non-local Means-based Denoising Algorithm in Brain CT Phantom Image using 3D Printing

Seong-Hyeon Kang, Youngjin Lee, Minji Park, Se‐Won Lim et al.
Daehan CT yeongsang gisul hakoeji
Medical Imaging and Analysis
article

A Study on Optimizing the Smoothing Factor for Fast Non-local Means-based Denoising Algorithm in Brain CT Phantom Image using 3D Printing

Seong-Hyeon Kang, Youngjin Lee, Minji Park, Se‐Won Lim, Hajin Kim
article en

Abstract

The purpose of this study is to fabricate a 3D printed phantom for computed tomography (CT) imaging and to optimize the smoothing factor of the fast non-local means (FNLM) denoising algorithm. For this purpose, a 3D printer with fused filament fabrication technology was used to produce the brain CT phantom with a diameter of 16 cm that mimics cerebrospinal fluid, gray matter, white matter, and bone. Then, we added the Gaussian noise of 0.005 to express the low-dose conditions from CT images acquired using a brain phantom. The optimization of performed by applying the FNLM algorithm to the noise-added CT phantom image. The d-value for adjusting smoothing intensity of FNLM algorithm was applied from 0.01 to 1.00 in increments of 0.01. The contrast to noise ratio (CNR), signal to noise ratio (SNR), and pixel profiles were measured for quantitative evaluation. As a result, both CNR and SNR were greatly improved until the d-value was 0.07, however, gradual changes were observed showing a flattened slope after 0.07. In addition, we confirmed that the signal value was stabilized at 0.07 in the measurement of the pixel profile. The incline exhibited a milder at subsequent d-values and the contrast was decreased. Based on these results, the smoothing factor of the FNLM algorithm was optimized to 0.07. Additionally, a comparative evaluation with the Wiener filter, median modified Wiener filter and total variation algorithm was conducted. As a result, the most improved CNR and SNR values were obtained in all tissues when applying the FNLM algorithm with an optimized d-value of 0.07. In conclusion, the FNLM algorithm with d-value of 0.07 was verified to be effective in reducing noise for low-dose CT conditions.

Daehan CT yeongsang gisul hakoejiVol. 26(2)
Openalex Percentile: Top 83%
Medical Imaging and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A Study on Optimizing the Smoothing Factor for Fast Non-local Means-based Denoising Algorithm in Brain CT Phantom Image using 3D Printing — Seong-Hyeon Kang, Youngjin Lee, et al. · Daehan CT yeongsang gisul hakoeji (2024) | TGRS Research Map | TGRS