Agreement in anterior segment measurements between swept-source optical coherence and dual Scheimpflug tomography devices in keratoconus eyes
The agreement between the two devices was not excellent for most parameters used to diagnose keratoconus and assess disease progression, and the differences were clinically significant. Therefore, the measurements from these two devices are not interchangeable for patients with keratoconus.
Authors
- Joo Youn Oh (ORCID: https://orcid.org/0000-0002-7899-5314)
- Hyuk Jin Choi (ORCID: https://orcid.org/0000-0002-7314-2267)
- Chang Ho Yoon (ORCID: https://orcid.org/0000-0003-2203-5709)
- Mee Kum Kim (ORCID: https://orcid.org/0000-0001-7850-6894)
- Yunjin Lee (ORCID: https://orcid.org/0000-0002-5627-8314)
Institutions
- Seoul National University (KR)
- New Generation University College (ET)
- Seoul National University Hospital (KR)
- National University College (PR)
- Deleted Institution
Publication Details
- Journal
- Scientific Reports
- Published
- 2024-10-17
- DOI
- https://doi.org/10.1038/s41598-024-76031-7
- Citations
- 3
- Primary Topic
- Corneal surgery and disorders
- Type
- article
- Field-Weighted Citation Impact
- 1.21