Study on the impact of bowtie filter on photon-counting CT imaging

<i>Objective.</i>The aim of this study was to investigate the impact of the bowtie filter on the image quality of the photon-counting detector (PCD) based CT imaging.<i>Approach.</i>Numerical simulations were conducted to investigate the impact of bowtie filters on image uniformity using two water phantoms, with tube potentials ranging from 60 to 140 kVp with a step of 5 kVp. Subsequently, benchtop PCD-CT imaging experiments were performed to verify the observations from the numerical simulations. Additionally, various correction methods were validated through these experiments.<i>Main results.</i>It was found that the use of a bowtie filter significantly alters the uniformity of PCD-CT images, depending on the size of the object and the x-ray spectrum. Two notable effects were observed: the capping effect and the flattening effect. Furthermore, it was demonstrated that the conventional beam hardening correction method could effectively mitigate such non-uniformity in PCD-CT images, provided that dedicated calibration parameters were used.<i>Significance.</i>It was demonstrated that the incorporation of a bowtie filter results in varied image artifacts in PCD-CT imaging under different conditions. Certain image correction methods can effectively mitigate and reduce these artifacts, thereby enhancing the overall quality of PCD-CT images.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2024-10-17
DOI
https://doi.org/10.1088/1361-6560/ad8858
Citations
1
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.18

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Study on the impact of bowtie filter on photon-counting CT imaging

Yongshuai Ge, Ting Su, Xin Zhang, Hairong Zheng et al.
1 citations
Physics in Medicine and Biology
Advanced X-ray and CT Imaging
0.18
article

Study on the impact of bowtie filter on photon-counting CT imaging

Yongshuai Ge, Ting Su, Xin Zhang, Hairong Zheng, Jiongtao Zhu, Dongmei Xia, Dong Liang, Jixiong Xie
article en
1 citations

Abstract

<i>Objective.</i>The aim of this study was to investigate the impact of the bowtie filter on the image quality of the photon-counting detector (PCD) based CT imaging.<i>Approach.</i>Numerical simulations were conducted to investigate the impact of bowtie filters on image uniformity using two water phantoms, with tube potentials ranging from 60 to 140 kVp with a step of 5 kVp. Subsequently, benchtop PCD-CT imaging experiments were performed to verify the observations from the numerical simulations. Additionally, various correction methods were validated through these experiments.<i>Main results.</i>It was found that the use of a bowtie filter significantly alters the uniformity of PCD-CT images, depending on the size of the object and the x-ray spectrum. Two notable effects were observed: the capping effect and the flattening effect. Furthermore, it was demonstrated that the conventional beam hardening correction method could effectively mitigate such non-uniformity in PCD-CT images, provided that dedicated calibration parameters were used.<i>Significance.</i>It was demonstrated that the incorporation of a bowtie filter results in varied image artifacts in PCD-CT imaging under different conditions. Certain image correction methods can effectively mitigate and reduce these artifacts, thereby enhancing the overall quality of PCD-CT images.

Physics in Medicine and BiologyVol. 69(21)
Chongqing University (CN), Chinese Academy of Sciences (CN), Key Laboratory of Guangdong Province (CN), Shenzhen Institutes of Advanced Technology (CN), University of Chinese Academy of Sciences (CN)
National Natural Science Foundation of China, Youth Innovation Promotion Association of the Chinese Academy of Sciences
Openalex Percentile: Top 52%
Advanced X-ray and CT Imaging
0.18
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.