Effect of Battery Level During Successive Charging Cycles on the Performance of Certified and Low-cost Uncertified Light-curing Units Available on E-commerce
The certified LCUs (VALO Grand, Radii Xpert, and LED.B) and uncertified LCU (VAFU), maintained power, emission spectrum, and light distributions during the tested battery life cycles. Low-cost certified LCU LED.B exhibited inhomogeneous light concentrated at the center of the tip. Low-cost uncertified LCUs-BS 300, 1 Sec, and LED curing light-had significant power reductions during the battery cycles and increased inhomogeneous light distribution along the successive exposure times.
Authors
- Carlos José Soares (ORCID: https://orcid.org/0000-0002-8830-605X)
- Margarida Faria
- Guilherme Saramago de Oliveira (ORCID: https://orcid.org/0000-0001-6638-7621)
- HL Carlo
- TS Peres
- SP da Silva Sakamoto
Institutions
- Universidade de Rio Verde (BR)
- Universidade Federal de Uberlândia (BR)
Publication Details
- Journal
- Operative Dentistry
- Published
- 2024-10-16
- DOI
- https://doi.org/10.2341/23-177-l
- Primary Topic
- Ergonomics and Musculoskeletal Disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.00