Long-Term Reproducibility of BMD-Measurements with Clinical QCT Using Simultaneous and Asynchronous Calibration Methods and Different Measurement and Reconstruction Protocols

Osteoporosis is underdiagnosed and undertreated. To improve timely fracture risk assessment optimized densitometry methods are required such as opportunistic spinal quantitative computed tomography (QCT). However, it is unclear how to best calibrate these scans and correct for potential scanner drift of QCT when used for monitoring bone mineral density (BMD) changes. We compared gold standard simultaneous calibration with asynchronous calibration methods, assessing mid-term (12 weeks) and long-term (1.5 years) reproducibility of BMD measurements. Cortical and trabecular compartments of the European Spine Phantom were studied with ten different protocols including low dose and high resolution (HR)-modes. Based on weekly phantom data, we compared simultaneous calibration to asynchronous single (termed global) or monthly calibration. The accuracy was better for trabecular measurements than for cortical measurements for all calibration methods. Reproducibility was excellent for all methods and slightly better for asynchronous than for simultaneous calibration both for trabecular and cortical bone. For HR protocols, reproducibility was better than for low dose measurements. In trabecular compartments averaged HR-BMD remained stable for global (- 0.1%/year, ns) but not for simultaneous calibration (- 1.5%/year, p < 0.001). No significant drifts could be detected for averaged low dose BMD (- 0.9 to + 0.8%/year) for either calibration method. Our data suggest that with regard to precision and accuracy measurements with asynchronous calibration are suitable for vertebral BMD assessment (no contrast agents) in clinical practice. Regular (e.g., monthly) stability tests using a calibration phantom could assure long term stability of at least 1 year.

Authors

Institutions

Publication Details

Journal
Calcified Tissue International
Published
2024-10-16
DOI
https://doi.org/10.1007/s00223-024-01303-3
Citations
6
Primary Topic
Bone health and osteoporosis research
Type
article
Field-Weighted Citation Impact
2.87

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Long-Term Reproducibility of BMD-Measurements with Clinical QCT Using Simultaneous and Asynchronous Calibration Methods and Different Measurement and Reconstruction Protocols

Marcus Both, Timo Damm, Claus‐Christian Glüer, Stefan Reinhold et al.
6 citations
Calcified Tissue International
Bone health and osteoporosis research
2.87
article

Long-Term Reproducibility of BMD-Measurements with Clinical QCT Using Simultaneous and Asynchronous Calibration Methods and Different Measurement and Reconstruction Protocols

Marcus Both, Timo Damm, Claus‐Christian Glüer, Stefan Reinhold, Jaime Peña, Reinhard Barkmann, Sophie du Mont, Meike Mainusch
article en
6 citations

Abstract

Osteoporosis is underdiagnosed and undertreated. To improve timely fracture risk assessment optimized densitometry methods are required such as opportunistic spinal quantitative computed tomography (QCT). However, it is unclear how to best calibrate these scans and correct for potential scanner drift of QCT when used for monitoring bone mineral density (BMD) changes. We compared gold standard simultaneous calibration with asynchronous calibration methods, assessing mid-term (12 weeks) and long-term (1.5 years) reproducibility of BMD measurements. Cortical and trabecular compartments of the European Spine Phantom were studied with ten different protocols including low dose and high resolution (HR)-modes. Based on weekly phantom data, we compared simultaneous calibration to asynchronous single (termed global) or monthly calibration. The accuracy was better for trabecular measurements than for cortical measurements for all calibration methods. Reproducibility was excellent for all methods and slightly better for asynchronous than for simultaneous calibration both for trabecular and cortical bone. For HR protocols, reproducibility was better than for low dose measurements. In trabecular compartments averaged HR-BMD remained stable for global (- 0.1%/year, ns) but not for simultaneous calibration (- 1.5%/year, p < 0.001). No significant drifts could be detected for averaged low dose BMD (- 0.9 to + 0.8%/year) for either calibration method. Our data suggest that with regard to precision and accuracy measurements with asynchronous calibration are suitable for vertebral BMD assessment (no contrast agents) in clinical practice. Regular (e.g., monthly) stability tests using a calibration phantom could assure long term stability of at least 1 year.

Calcified Tissue InternationalVol. 115(5)
Christian-Albrechts-Universität zu Kiel (DE), University Hospital Schleswig-Holstein (DE), University of Lübeck (DE)
Bundesministerium für Bildung und Forschung
Openalex Percentile: Top 9%
Bone health and osteoporosis research
2.87
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.