High Resolving Power Electrospray Ionization Ion Mobility Spectrometer Based on Fourier Deconvolution Multiplexing

The resolving power of the drift tube ion mobility spectrometry (IMS) is mainly dependent on the drift length, the drift voltage, the pulse width of an ion gate, and the pressure inside the drift tube. Electrospray ionization (ESI)-IMS is a highly sensitive and reliable technique for the detection and analysis of nonvolatile compounds, and high resolving power is necessary to separate structurally similar compounds. To improve the analytical performance of atmospheric pressure ESI-IMS, the Fourier deconvolution (FD) multiplexing technique is investigated as an effective and convenient means to improve the resolving power as well as the signal-to-noise ratio. By reducing the equivalent ion gate opening width to 5 μs using a typical Tyndall-Powell ion shutter, a high resolving power <i>R</i><sub>P</sub> up to 170 can be achieved with a drift length of 12 cm and a drift voltage of 10 kV. Rhodamine 6G (R6G), sodium dodecyl sulfate (SDS), methacycline, oxytetracycline, and ractopamine were evaluated using the FD-ESI-IMS, and mixtures with similar ion mobility can be well separated without prolonging the drift length.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2024-10-16
DOI
https://doi.org/10.1021/acs.analchem.4c04549
Citations
7
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.69

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

High Resolving Power Electrospray Ionization Ion Mobility Spectrometer Based on Fourier Deconvolution Multiplexing

Jianna Yu, Feiyu Lu, Jiaxu Liu, Lian Duan et al.
7 citations
Analytical Chemistry
Mass Spectrometry Techniques and Applications
1.69
article

High Resolving Power Electrospray Ionization Ion Mobility Spectrometer Based on Fourier Deconvolution Multiplexing

Jianna Yu, Feiyu Lu, Jiaxu Liu, Lian Duan, Wenshan Li, Binwang Yang, Guoxing Jing, Wenjie Liu, Wen Liu, Zhijian Yang, Xue Zheng, Zhixiong Ye, Cheng Wang
article en
7 citations

Abstract

The resolving power of the drift tube ion mobility spectrometry (IMS) is mainly dependent on the drift length, the drift voltage, the pulse width of an ion gate, and the pressure inside the drift tube. Electrospray ionization (ESI)-IMS is a highly sensitive and reliable technique for the detection and analysis of nonvolatile compounds, and high resolving power is necessary to separate structurally similar compounds. To improve the analytical performance of atmospheric pressure ESI-IMS, the Fourier deconvolution (FD) multiplexing technique is investigated as an effective and convenient means to improve the resolving power as well as the signal-to-noise ratio. By reducing the equivalent ion gate opening width to 5 μs using a typical Tyndall-Powell ion shutter, a high resolving power <i>R</i><sub>P</sub> up to 170 can be achieved with a drift length of 12 cm and a drift voltage of 10 kV. Rhodamine 6G (R6G), sodium dodecyl sulfate (SDS), methacycline, oxytetracycline, and ractopamine were evaluated using the FD-ESI-IMS, and mixtures with similar ion mobility can be well separated without prolonging the drift length.

Analytical ChemistryVol. 96(43)
Xiangtan University (CN)
National Natural Science Foundation of China
Affordable and clean energy
Openalex Percentile: Top 16%
Mass Spectrometry Techniques and Applications
1.69
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.