Conservation of the Charge in Signal from Drift Tube Ion Mobility Spectrometers

Quantitative information obtained from drift tube detectors used in ion mobility spectrometry is contained in the area of peaks forming the drift time spectrum. The area of all peaks corresponds to the total charge of ions entering the drift section of the spectrometer. It was found that this charge is not conserved when the ion composition changes. This work is devoted to studying the causes of this phenomenon. Experimental research consisted of recording drift time spectra for 2-pentanone and <i>n</i>-heptanone, at various analyte concentrations and different opening times of the shutter grid. Measurements of the total ion current were also performed in static mode with an open grid. The research results indicated that the reasons for the lack of ion charge conservation in the drift time spectrum are ion recombination, mutual repulsion, and mobility-dependent transmission of ions through the shutter grid. The explanation of the relationships obtained experimentally was based on a simple theoretical model, which considered the phenomenon of ion transport along the reaction section and the penetration of ions through the shutter. The developed model provides a good description of the measurement results and allows the estimation of ion currents and ion concentrations in the reaction section upstream of the grid. This information is important for proper quantitative analysis as well as when the ion mobility spectrometer is used in quantitative studies of chemical ionization processes.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2024-10-16
DOI
https://doi.org/10.1021/acs.analchem.4c03825
Citations
3
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.72

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Conservation of the Charge in Signal from Drift Tube Ion Mobility Spectrometers

Jarosław Puton, Izabela Wolańska, Krzysztof Piwowarski
3 citations
Analytical Chemistry
Mass Spectrometry Techniques and Applications
0.72
article

Conservation of the Charge in Signal from Drift Tube Ion Mobility Spectrometers

Jarosław Puton, Izabela Wolańska, Krzysztof Piwowarski
article en
3 citations

Abstract

Quantitative information obtained from drift tube detectors used in ion mobility spectrometry is contained in the area of peaks forming the drift time spectrum. The area of all peaks corresponds to the total charge of ions entering the drift section of the spectrometer. It was found that this charge is not conserved when the ion composition changes. This work is devoted to studying the causes of this phenomenon. Experimental research consisted of recording drift time spectra for 2-pentanone and <i>n</i>-heptanone, at various analyte concentrations and different opening times of the shutter grid. Measurements of the total ion current were also performed in static mode with an open grid. The research results indicated that the reasons for the lack of ion charge conservation in the drift time spectrum are ion recombination, mutual repulsion, and mobility-dependent transmission of ions through the shutter grid. The explanation of the relationships obtained experimentally was based on a simple theoretical model, which considered the phenomenon of ion transport along the reaction section and the penetration of ions through the shutter. The developed model provides a good description of the measurement results and allows the estimation of ion currents and ion concentrations in the reaction section upstream of the grid. This information is important for proper quantitative analysis as well as when the ion mobility spectrometer is used in quantitative studies of chemical ionization processes.

Analytical ChemistryVol. 96(43)
Military University of Technology in Warsaw (PL)
Narodowe Centrum Badań i Rozwoju
Openalex Percentile: Top 33%
Mass Spectrometry Techniques and Applications
0.72
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Conservation of the Charge in Signal from Drift Tube Ion Mobility Spectrometers — Jarosław Puton, Izabela Wolańska, et al. · Analytical Chemistry (2024) | TGRS Research Map | TGRS