Performance of a BeO-based dosimetry system for proton and electron beam dose measurements

This study aims to evaluate the performance of the BeO-based myOSLchip system (RadPro International GmbH, Remscheid, Germany) for dosimetry of proton and electron radiotherapy beams. Although beryllium oxide (BeO) has been recognised as a promising material for luminescence dosimetry in radiotherapy, this research extends beyond material properties and examines the entire BeO-based dosimetry system, including the detector, holder, and readout components. Packages of myOSLchip dosimeters were irradiated either in a ( 70 − 230 ) MeV proton beam or in a 16 MeV electron beam. The readouts were carried out using the portable myOSLchip reader. In the electron beam, tests on the precision, dose response up to 100 Gy and dose-rate effects of the system were carried out. In the proton beam, the system was tested for its dose response (up to 10 Gy), fading, and linear energy transfer (LET) response. For proton irradiations, the myOSLchip BeO OSLDs exhibited stability within 2% over 135 days, as well as a linear dose response in the tested range, ( 0 . 1 − 10 ) Gy. The efficiency showed a reduction for proton beams with LET values (for water) above 0.6 keV/ μ m, with up to 40% loss in efficiency at 4 keV/ μ m. For the electron irradiations, they showed a linear dose–response up to 20 Gy and dose-rate independence, with a constant response at least up to 2.99 × 10 5 Gy s −1 . Using individual dosimeter sensitivity correction, the precision for a single dosimeter was around 3.5% (standard deviation of the data of all dosimeters) and for a package comprising four dosimeters was 1.7% (standard deviation of the mean of the four dosimeters). These findings suggest the myOSLchip system’s potential as a reliable alternative to existing dosimetry systems in clinical applications. • A single-detector BeO-based dosimetry system was investigated in proton and electron beams. • Performances tested include precision, dose response, dose-rate effects, fading and LET response. • A reduced efficiency was observed at LET values > 0.6 keV/ μ m. • The performances indicated the system’s capability to be used for clinical applications.

Authors

Institutions

Publication Details

Journal
Radiation Measurements
Published
2024-10-16
DOI
https://doi.org/10.1016/j.radmeas.2024.107310
Citations
5
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
2.09
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance of a BeO-based dosimetry system for proton and electron beam dose measurements

Jeppe Brage Christensen, Riccardo Dal Bello, E.G. Yukihara, Silvia Motta et al.
5 citations
Radiation Measurements
Radiation Therapy and Dosimetry
2.09
article

Performance of a BeO-based dosimetry system for proton and electron beam dose measurements

Jeppe Brage Christensen, Riccardo Dal Bello, E.G. Yukihara, Silvia Motta, Michele Togno, Lily Bossin, Stefan Schischke
article en
5 citations

Abstract

This study aims to evaluate the performance of the BeO-based myOSLchip system (RadPro International GmbH, Remscheid, Germany) for dosimetry of proton and electron radiotherapy beams. Although beryllium oxide (BeO) has been recognised as a promising material for luminescence dosimetry in radiotherapy, this research extends beyond material properties and examines the entire BeO-based dosimetry system, including the detector, holder, and readout components. Packages of myOSLchip dosimeters were irradiated either in a ( 70 − 230 ) MeV proton beam or in a 16 MeV electron beam. The readouts were carried out using the portable myOSLchip reader. In the electron beam, tests on the precision, dose response up to 100 Gy and dose-rate effects of the system were carried out. In the proton beam, the system was tested for its dose response (up to 10 Gy), fading, and linear energy transfer (LET) response. For proton irradiations, the myOSLchip BeO OSLDs exhibited stability within 2% over 135 days, as well as a linear dose response in the tested range, ( 0 . 1 − 10 ) Gy. The efficiency showed a reduction for proton beams with LET values (for water) above 0.6 keV/ μ m, with up to 40% loss in efficiency at 4 keV/ μ m. For the electron irradiations, they showed a linear dose–response up to 20 Gy and dose-rate independence, with a constant response at least up to 2.99 × 10 5 Gy s −1 . Using individual dosimeter sensitivity correction, the precision for a single dosimeter was around 3.5% (standard deviation of the data of all dosimeters) and for a package comprising four dosimeters was 1.7% (standard deviation of the mean of the four dosimeters). These findings suggest the myOSLchip system’s potential as a reliable alternative to existing dosimetry systems in clinical applications. • A single-detector BeO-based dosimetry system was investigated in proton and electron beams. • Performances tested include precision, dose response, dose-rate effects, fading and LET response. • A reduced efficiency was observed at LET values > 0.6 keV/ μ m. • The performances indicated the system’s capability to be used for clinical applications.

Radiation MeasurementsVol. 179
University of Zurich (CH), Paul Scherrer Institute (CH), University Hospital of Zurich (CH), Bio-Rad (Germany) (DE)
Affordable and clean energy
Openalex Percentile: Top 12%
Radiation Therapy and Dosimetry
2.09
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.