Experimental Assessment of Magnetic Resonance Imaging Distortionfor Radiation Therapy Planning

MRI is widely used in radiation therapy planning, particularly in stereotactic radiosurgery for brain metastases. The article addresses the geometric distortion of MRI images, which can lead to errors in radiation therapy planning. A series of experiments with a simple phantom were conducted on two MRI scanners with 0.5 T and 1.5 T magnetic fields. Deviations in the positions of the phantom objects from their actual locations were observed. The detected deviations can reach up to 5 mm. A theoretical dependence of the magnetic field gradient on the distance to the center of homogeneity was calculated, which agrees well with the approximation of experimental data. An assessment was made of the dependence of the image area distortion of objects located at the center of the field homogeneity on their actual sizes. It was found that the area deviation can reach up to 20%.

Authors

Institutions

Publication Details

Journal
Vestnik Moskovskogo Universiteta Seriya 3 Fizika Astronomiya
Published
2024-10-14
DOI
https://doi.org/10.55959/msu0579-9392.79.2440703
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Experimental Assessment of Magnetic Resonance Imaging Distortionfor Radiation Therapy Planning

С. А. Гаврилова, Е. Н. Лыкова, A. P. Chernyaev, Ф. Р. Студеникин et al.
Vestnik Moskovskogo Universiteta Seriya 3 Fizika Astronomiya
Advanced MRI Techniques and Applications
article

Experimental Assessment of Magnetic Resonance Imaging Distortionfor Radiation Therapy Planning

С. А. Гаврилова, Е. Н. Лыкова, A. P. Chernyaev, Ф. Р. Студеникин, A. A. Shcherbakov, A. P. Strelkovskaya
article en

Abstract

MRI is widely used in radiation therapy planning, particularly in stereotactic radiosurgery for brain metastases. The article addresses the geometric distortion of MRI images, which can lead to errors in radiation therapy planning. A series of experiments with a simple phantom were conducted on two MRI scanners with 0.5 T and 1.5 T magnetic fields. Deviations in the positions of the phantom objects from their actual locations were observed. The detected deviations can reach up to 5 mm. A theoretical dependence of the magnetic field gradient on the distance to the center of homogeneity was calculated, which agrees well with the approximation of experimental data. An assessment was made of the dependence of the image area distortion of objects located at the center of the field homogeneity on their actual sizes. It was found that the area deviation can reach up to 20%.

Vestnik Moskovskogo Universiteta Seriya 3 Fizika Astronomiya(№4_2024)
Lomonosov Moscow State University (RU), Moscow State University (TJ)
Lomonosov Moscow State University
Openalex Percentile: Top 70%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Experimental Assessment of Magnetic Resonance Imaging Distortionfor Radiation Therapy Planning — С. А. Гаврилова, Е. Н. Лыкова, et al. · Vestnik Moskovskogo Universiteta Seriya 3 Fizika Astronomiya (2024) | TGRS Research Map | TGRS