Diagnostic accuracy of a non-invasive spot-check hemoglobin meter, Masimo Rad-67® pulse CO-Oximeter®, in detection of anemia in antenatal care settings in Kenya
Overall, SpHb by Masimo Rad-67® Pulse CO-Oximeter® did not accurately identify pregnant women with anemia and many cases would be missed. We would not recommend its use in antenatal care settings.
Authors
- Grace Mkanjala Mwashigadi (ORCID: https://orcid.org/0000-0002-6321-3287)
- Hiten D. Mistry (ORCID: https://orcid.org/0000-0003-2564-7348)
- Stanley Lüchters (ORCID: https://orcid.org/0000-0001-5235-5629)
- Marleen Temmerman (ORCID: https://orcid.org/0000-0003-2069-8752)
- Angela C. Koech (ORCID: https://orcid.org/0000-0003-4838-3269)
- Geoffrey Omuse (ORCID: https://orcid.org/0000-0002-3866-5118)
- Peter von Dadelszen (ORCID: https://orcid.org/0000-0003-4136-3070)
- Isaac G. Mwaniki (ORCID: https://orcid.org/0009-0003-1299-1657)
- Moses W. Mukhanya (ORCID: https://orcid.org/0000-0002-3954-9426)
- Joseph Mutunga
- Marvine Ochieng
- Rachel Craik
- Emily Mwadime
Institutions
- King's College London (GB)
- Liverpool School of Tropical Medicine (GB)
- Ghent University (BE)
- Aga Khan University Nairobi (KE)
- Centre for Sexual Health and HIV AIDS Research (ZW)
Publication Details
- Journal
- Frontiers in Global Women s Health
- Published
- 2024-10-14
- DOI
- https://doi.org/10.3389/fgwh.2024.1427261
- Citations
- 8
- Primary Topic
- Iron Metabolism and Disorders
- Type
- article
- Field-Weighted Citation Impact
- 3.64