Comparison of ocular biometric measurements and intraocular lens power calculation using different methods in eyes with implantable collamer lenses

This prospective cohort study included 80 healthy candidates for Implantable Collamer Lens (ICL) implantation who underwent biometric assessments with Scheimpflug imaging (the Pentacam-AXL) and swept-source optical coherence tomography (SS-OCT; the IOLMaster-700), both before and 3 months after surgery. The main outcome measures were mean keratometry, anterior chamber depth, axial length, and various intraocular lens (IOL) calculation formulas (Haigis, SRK/T, Hoffer Q, Holladay 1, Barrett Universal 2, and Olsen). The interchangeability of the devices was assessed by generating 95% limits of agreement (95% LoA) and associated Bland-Altman plots. The average age of the participants was 31.5 ± 5.4 years (22–43), with 58 (72.5%) being female. Among the cases analyzed, 11 (13.4%) had incorrect anterior lens surface segmentation using the IOLMaster-700, and 1 case (1.2%) had inappropriate segmentation using the Pentacam-AXL. Postoperative IOL power calculation resulted in readings that were, on average, 0.15 to 0.30 D higher compared to preoperative measurements. The 95% LoAs could differ by up to 0.85 D higher after surgery, indicating weak agreement between pre- and postoperative measurements. There was poor agreement between the IOLMaster-700 and Pentacam-AXL in IOL power calculation for eyes with post-ICL implantation, with a difference of more than 1 D in the 95% LoAs. In conclusion, Scheimpflug imaging was found to be less susceptible than the SS-OCT technique to segmentation errors of the anterior lens surface after ICL implantation. Neither device showed interchangeable results for pre- versus postoperative IOL power calculation. The determination of IOL power by the IOLMaster-700 versus Pentacam-AXL was not interchangeable in eyes with ICL implantation.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-10-14
DOI
https://doi.org/10.1038/s41598-024-74880-w
Citations
3
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
1.16
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of ocular biometric measurements and intraocular lens power calculation using different methods in eyes with implantable collamer lenses

Kia Salouti, Maryam Ghoreyshi, M. Hossein Nowroozzadeh, Ramin Salouti et al.
3 citations
Scientific Reports
Ophthalmology and Visual Impairment Studies
1.16
article

Comparison of ocular biometric measurements and intraocular lens power calculation using different methods in eyes with implantable collamer lenses

Kia Salouti, Maryam Ghoreyshi, M. Hossein Nowroozzadeh, Ramin Salouti, Mostafa Nazarpour-Servak
article en
3 citations

Abstract

This prospective cohort study included 80 healthy candidates for Implantable Collamer Lens (ICL) implantation who underwent biometric assessments with Scheimpflug imaging (the Pentacam-AXL) and swept-source optical coherence tomography (SS-OCT; the IOLMaster-700), both before and 3 months after surgery. The main outcome measures were mean keratometry, anterior chamber depth, axial length, and various intraocular lens (IOL) calculation formulas (Haigis, SRK/T, Hoffer Q, Holladay 1, Barrett Universal 2, and Olsen). The interchangeability of the devices was assessed by generating 95% limits of agreement (95% LoA) and associated Bland-Altman plots. The average age of the participants was 31.5 ± 5.4 years (22–43), with 58 (72.5%) being female. Among the cases analyzed, 11 (13.4%) had incorrect anterior lens surface segmentation using the IOLMaster-700, and 1 case (1.2%) had inappropriate segmentation using the Pentacam-AXL. Postoperative IOL power calculation resulted in readings that were, on average, 0.15 to 0.30 D higher compared to preoperative measurements. The 95% LoAs could differ by up to 0.85 D higher after surgery, indicating weak agreement between pre- and postoperative measurements. There was poor agreement between the IOLMaster-700 and Pentacam-AXL in IOL power calculation for eyes with post-ICL implantation, with a difference of more than 1 D in the 95% LoAs. In conclusion, Scheimpflug imaging was found to be less susceptible than the SS-OCT technique to segmentation errors of the anterior lens surface after ICL implantation. Neither device showed interchangeable results for pre- versus postoperative IOL power calculation. The determination of IOL power by the IOLMaster-700 versus Pentacam-AXL was not interchangeable in eyes with ICL implantation.

Scientific ReportsVol. 14(1)
University of British Columbia (CA), Shiraz University of Medical Sciences (IR)
Openalex Percentile: Top 19%
Ophthalmology and Visual Impairment Studies
1.16
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.