Validity and Reliability of a New Wearable Chest Strap to Estimate Respiratory Frequency in Elite Soccer Athletes
Assessing respiratory frequency (<i>f</i><sub>R</sub>) is practical in monitoring training progress in competitive athletes, especially during exercise. This study aimed to validate a new wearable chest strap (wCS) to estimate <i>f</i><sub>R</sub> against ergospirometry as a criterion device in soccer players. A total of 26 elite professional soccer players (mean [standard deviation]: 23.6 [4.8] years; 180.6 [5.7] cm; 77.2 [5.4] kg) from three Italian Serie A League teams participated in this cross-sectional study. The sample included attackers, midfielders, and defenders. <i>f</i><sub>R</sub> was assessed during a maximal cardiopulmonary exercise test (CPET) on a treadmill using (i) a breath-by-breath gas exchange analyzer (Vyntus<sup>®</sup> CPX, Vyaire Medical) and (ii) a novel wCS with sensors designed to assess breath frequency following chest expansions. Pearson's correlation coefficient (<i>r</i>), adjusted coefficient of determination (aR<sup>2</sup>), Bland-Altman plot analysis, and Lin's concordance correlation coefficient (ρ<sub>c</sub>) were used for comparative analysis (correlation and concordance) among the methods. The repeated measures correlation coefficient (r<sub>rm</sub>) was used to assess the strength of the linear association between the methods. The intraclass correlation coefficient (ICC) and the Finn coefficient (r<sub>F</sub>) were used for inter-rater reliability. All statistical analyses were performed within the R statistical computing environment, with 95% confidence intervals (95% CIs) reported and statistical significance set at <i>p</i> < 0.05. A total of 16529 comparisons were performed after collecting the CPET data. The robust time series analysis with Hodges-Lehmann estimation showed no significant differences between both methods (<i>p</i> > 0.05). Correlation among devices was statistically significant and very large (<i>r</i> [95% CI]: 0.970 [0.970, 0.971], <i>p</i> < 0.01; aR<sup>2</sup> [95% CI]: 0.942 [0.942, 0.943], <i>p</i> < 0.01) with strong evidence supporting consistency of the new wCS (BF<sub>10</sub> > 100). In addition, a high concordance was found (ρ<sub>c</sub> [95% CI]: 0.970 [0.969, 0.971], bias correction factor: 0.999). Vyntus<sup>TM</sup> CPX, as a standard criterion, showed moderate agreement with wCS after Bland-Altman analysis (bias [95% lower to the upper limit of agreement]; % agree: 0.170 [-4.582 to 4.923] breaths·min<sup>-1</sup>; 69.9%). A strong association between measurements (r<sub>rm</sub> [95% CI]: 0.960 [0.959, 0.961]), a high absolute agreement between methods (ICC [95% CI]: 0.970 [0.970, 0.971]), and high inter-rater reliability (r<sub>F</sub>: 0.947) were found. With an RMSE = 2.42 breaths·min<sup>-1</sup>, the new wCS seems to be an valid and reliable in-field method to evaluate <i>f</i><sub>R</sub> compared to a breath-by-breath gas exchange analyzer. Notwithstanding, caution is advised if methods are used interchangeably while further external validation occurs.
Authors
- Erika Cione (ORCID: https://orcid.org/0000-0002-0562-0597)
- Diego A. Bonilla (ORCID: https://orcid.org/0000-0002-2634-1220)
- Roberto Cannataro (ORCID: https://orcid.org/0000-0002-1668-7690)
- Adriano Di Paco
- Raffaele Canonico
- Rocco Perrotta
Institutions
- University of the Basque Country (ES)
- University of Campania "Luigi Vanvitelli" (IT)
- Azienda Unità Sanitaria Locale 11 di Empoli (IT)
- University of Calabria (IT)
- Universidad Católica de Murcia (ES)
Publication Details
- Journal
- Sports
- Published
- 2024-10-12
- DOI
- https://doi.org/10.3390/sports12100277
- Citations
- 2
- Primary Topic
- Cardiovascular and exercise physiology
- Type
- article
- Field-Weighted Citation Impact
- 1.43
Funders
- Università della Calabria