Testing the accuracy of low-beam-energy electron-excited X-ray microanalysis with energy-dispersive spectrometry

The accuracy of electron-excited X-ray microanalysis with energy-dispersive spectrometry (EDS) has been tested in the low beam energy range, specifically at an incident beam energy of 5 keV, which is the lowest beam energy for which a useful characteristic X-ray peak can be excited for all elements of the periodic table, excepting H and He. Elemental analysis results are reported for certified reference materials (CRM), stoichiometric compounds, minerals, and metal alloys of independently known or measured composition which had microscopic homogeneity suitable for microanalysis. Two-hundred sixty-three concentration measurements for 39 elements in 113 materials were determined following the <i>k-ratio protocol</i> and using the EDS analytical software NIST DTSA-II. The accuracy of the results, as characterized by the <i>relative deviation from expected value</i> (RDEV) metric, was such that more than 98% of the results were found to be captured within a range of ±5% RDEV, while 82% of the results fell in the range -2% to 2% RDEV.

Authors

Institutions

Publication Details

Journal
Journal of Materials Science
Published
2024-10-01
DOI
https://doi.org/10.1007/s10853-024-10285-4
Citations
7
Primary Topic
Electron and X-Ray Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
1.25

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Testing the accuracy of low-beam-energy electron-excited X-ray microanalysis with energy-dispersive spectrometry

Dale E. Newbury, Nicholas W. M. Ritchie
7 citations
Journal of Materials Science
Electron and X-Ray Spectroscopy Techniques
1.25
article

Testing the accuracy of low-beam-energy electron-excited X-ray microanalysis with energy-dispersive spectrometry

Dale E. Newbury, Nicholas W. M. Ritchie
article en
7 citations

Abstract

The accuracy of electron-excited X-ray microanalysis with energy-dispersive spectrometry (EDS) has been tested in the low beam energy range, specifically at an incident beam energy of 5 keV, which is the lowest beam energy for which a useful characteristic X-ray peak can be excited for all elements of the periodic table, excepting H and He. Elemental analysis results are reported for certified reference materials (CRM), stoichiometric compounds, minerals, and metal alloys of independently known or measured composition which had microscopic homogeneity suitable for microanalysis. Two-hundred sixty-three concentration measurements for 39 elements in 113 materials were determined following the <i>k-ratio protocol</i> and using the EDS analytical software NIST DTSA-II. The accuracy of the results, as characterized by the <i>relative deviation from expected value</i> (RDEV) metric, was such that more than 98% of the results were found to be captured within a range of ±5% RDEV, while 82% of the results fell in the range -2% to 2% RDEV.

Journal of Materials ScienceVol. 59(40)
National Institute of Standards and Technology (US)
National Institute of Standards and Technology
Affordable and clean energy
Openalex Percentile: Top 23%
Electron and X-Ray Spectroscopy Techniques
1.25
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.