Bivalirudin anticoagulation for an infant with heparin resistance on ECMO: A case report
Due to the unique physiological characteristics of newborns, the hemostatic changes differ significantly from those in adults. Precise monitoring of anticoagulation becomes a critical and challenging task. Bivalirudin can be effectively used for anticoagulation management in neonatal ECMO; however, due to its unique characteristics, precise dose adjustment poses a challenge. Selecting the optimal laboratory monitoring indicators is crucial in this regard. In some cases, bedside ACT may not be the optimal anticoagulation monitoring parameter, and when necessary, comparative analysis can be conducted using anticoagulant-sample ACTs such as thrombela stogram-ACT. Traditional markers such as D-dimer/fibrinogen degradation products and newer indicators like TAT can reflect the activation of coagulation and assist in monitoring the anticoagulation effect, especially when there is conflicting information among the monitoring parameters.
Authors
- Lan Chen (ORCID: https://orcid.org/0000-0003-4520-5109)
- Siqi Guo (ORCID: https://orcid.org/0009-0005-1067-9330)
- Ge Zhang (ORCID: https://orcid.org/0009-0001-6136-6378)
- Jing Shi
Institutions
- Sichuan University (CN)
- West China Second University Hospital of Sichuan University (CN)
Publication Details
- Journal
- Medicine
- Published
- 2024-10-11
- DOI
- https://doi.org/10.1097/md.0000000000039357
- Citations
- 1
- Primary Topic
- Mechanical Circulatory Support Devices
- Type
- article
- Field-Weighted Citation Impact
- 0.18
Funders
- Sichuan University