New prefilled syringe aflibercept design. A cause of symptomatic IOP spike after aflibercept PFS?
Intraocular pressure spikes after aflibercept PFS can be explained by injector characteristics. The PFS of aflibercept has a domed plunger. Incorrect alignment between the base of the plunger and the black dosing line could cause an increase in the injected volume. Furthermore, the wider syringe diameter of aflibercept PFS could imply a larger injection force, increasing the risk of IOP elevation. Patient characteristics, such as previous VPP, axial length, or glaucoma history, may also play a role. Further studies are required to develop an ideal intravitreal syringe.
Authors
- Daniela Rego‐Lorca (ORCID: https://orcid.org/0000-0002-8076-0522)
- José I. Vela (ORCID: https://orcid.org/0000-0002-4310-5748)
- Jesús Díaz‐Cascajosa
- Irene Loscos‐Giménez
- Francisca Bassaganyas-Vilarrasa
- Jaume Crespí‐Vilimelis
Institutions
- Universitat Autònoma de Barcelona (ES)
- Hospital de Sant Pau (ES)
Publication Details
- Journal
- Romanian Journal of Ophthalmology
- Published
- 2024-10-07
- DOI
- https://doi.org/10.22336/rjo.2024.41
- Citations
- 1
- Primary Topic
- Health Systems, Economic Evaluations, Quality of Life
- Type
- article
- Field-Weighted Citation Impact
- 1.13