Towards high throughput analysis using 96-well plate solid phase extraction to determine sedatives and β-blocker residues in food control monitoring
This study is the first to present an optimized and validated holistic method for sedatives and β-blocker using 96-well plate SPE and UHPLC-MS/MS. The method showed good performance in kidney, meat and fish samples being an universal method for any species along the same type of sample. The fast, easy, efficient, reliable and universal method showed high throughput and so reduced the analysis time by nine-fold and the required solvent amount by four times fulfilling the green chemistry principals.
Authors
- Dieke van Doorn
- Ane Arrizabalaga-Larrañaga (ORCID: https://orcid.org/0000-0002-9379-5775)
- Saskia S. Sterk (ORCID: https://orcid.org/0000-0001-5225-1862)
Institutions
- Wageningen University & Research (NL)
Publication Details
- Journal
- Analytica Chimica Acta
- Published
- 2024-10-14
- DOI
- https://doi.org/10.1016/j.aca.2024.343335
- Citations
- 3
- Primary Topic
- Analytical chemistry methods development
- Type
- article
- Field-Weighted Citation Impact
- 0.71
Funders
- European Commission
- Ministerie van Landbouw, Natuur en Voedselkwaliteit
- Ministerie van Economische Zaken en Klimaat