A novel technique of airway silicon stent deployment under vision—Dr. Vidyasagar's technique

The conventional methods of silicon stent insertion recommend usage of external loading devices, where the stent is folded into the loading device and pushed in to the tracheobronchial tree using an external pusher which is blind, and leads to placement of stent either distally or proximally needing repositioning or is done with fluoroscopy that involves radiation exposure. We demonstrate our experience in 16 cases of successful silicon stent placement using this technique, wherein an Ultrathin flexible bronchoscope or Hopkins Rigid telescope is pushed alongside the forceps that hold upper end of the folded silicon stent allowing stent placement under direct vision with control over the stent. The Proximal end of the stent can be pulled under vision before deployment for appropriate positioning while pulling the rigid barrel. The stent is always under the operator's control providing excellent control over placement, simplifies the procedure and is safe with no reported complications.

Authors

Institutions

Publication Details

Journal
Respirology Case Reports
Published
2024-10-01
DOI
https://doi.org/10.1002/rcr2.70042
Citations
1
Primary Topic
Tracheal and airway disorders
Type
article
Field-Weighted Citation Impact
0.42
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A novel technique of airway silicon stent deployment under vision—Dr. Vidyasagar's technique

Sejal B. Radia, Hari Kishan Gonuguntla, Belgundi Preeti Vidyasagar, Suhas Dhulipala
1 citations
Respirology Case Reports
Tracheal and airway disorders
0.42
article

A novel technique of airway silicon stent deployment under vision—Dr. Vidyasagar's technique

Sejal B. Radia, Hari Kishan Gonuguntla, Belgundi Preeti Vidyasagar, Suhas Dhulipala
article en
1 citations

Abstract

The conventional methods of silicon stent insertion recommend usage of external loading devices, where the stent is folded into the loading device and pushed in to the tracheobronchial tree using an external pusher which is blind, and leads to placement of stent either distally or proximally needing repositioning or is done with fluoroscopy that involves radiation exposure. We demonstrate our experience in 16 cases of successful silicon stent placement using this technique, wherein an Ultrathin flexible bronchoscope or Hopkins Rigid telescope is pushed alongside the forceps that hold upper end of the folded silicon stent allowing stent placement under direct vision with control over the stent. The Proximal end of the stent can be pulled under vision before deployment for appropriate positioning while pulling the rigid barrel. The stent is always under the operator's control providing excellent control over placement, simplifies the procedure and is safe with no reported complications.

Respirology Case ReportsVol. 12(10)
Yashoda Hospital (IN)
Openalex Percentile: Top 32%
Tracheal and airway disorders
0.42
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A novel technique of airway silicon stent deployment under vision—Dr. Vidyasagar's technique — Sejal B. Radia, Hari Kishan Gonuguntla, et al. · Respirology Case Reports (2024) | TGRS Research Map | TGRS