Image quality of photon-counting detector CT for visualization of maxillofacial anatomy in comparison with energy-integrating detector CT and intraoperative C-arm CBCT
PCD-CT exhibits superior image quality over CBCT in key parameters essential for maxillofacial imaging, while no apparent improvement was shown compared to state-of-the-art EID-CT. PCD-CT offers enhanced visualization of critical anatomical structures, suggesting its potential as a preferred modality in managing maxillofacial trauma. The findings in this study align with limited existing research on PCD-CT, underscoring its promise for advanced diagnostic imaging in maxillofacial applications.
Authors
- Erik Tesselaar (ORCID: https://orcid.org/0000-0002-8387-0583)
- Mischa Woisetschläger (ORCID: https://orcid.org/0000-0003-0066-4985)
- Michael Sandborg (ORCID: https://orcid.org/0000-0003-3352-8330)
- Eva Klintström (ORCID: https://orcid.org/0000-0003-0884-899X)
- Adam Ly
Institutions
- Linköping University (SE)
Publication Details
- Journal
- European Journal of Radiology
- Published
- 2024-10-13
- DOI
- https://doi.org/10.1016/j.ejrad.2024.111785
- Citations
- 7
- Primary Topic
- Advanced X-ray and CT Imaging
- Type
- article
- Field-Weighted Citation Impact
- 1.29