Dynamic contrast-enhanced MRA of the aorta using a Golden-angle RAdial Sparse Parallel (GRASP) sequence: comparison with conventional time-resolved cartesian MRA (TWIST)
GRASP outperformed TWIST in SNR, vessel sharpness, and reduction in image blurring; streaking artefacts were stronger with GRASP, but did not affect diagnostic image quality.
Authors
- Bernd Jung (ORCID: https://orcid.org/0009-0002-3659-0930)
- Fabian Haupt (ORCID: https://orcid.org/0000-0002-2401-922X)
- Hendrik von Tengg‐Kobligk (ORCID: https://orcid.org/0000-0003-3207-3223)
- Alan A. Peters (ORCID: https://orcid.org/0000-0002-9864-6381)
- Adrian Thomas Huber (ORCID: https://orcid.org/0000-0001-6146-8238)
- Elena Kleban (ORCID: https://orcid.org/0000-0002-9814-0807)
- Fabrice Helfenstein (ORCID: https://orcid.org/0000-0001-8412-0461)
- Camilla Giulia Calastra (ORCID: https://orcid.org/0009-0003-7051-7056)
Institutions
- University of Bern (CH)
- University Hospital of Bern (CH)
Publication Details
- Journal
- The International Journal of Cardiovascular Imaging
- Published
- 2024-10-12
- DOI
- https://doi.org/10.1007/s10554-024-03259-9
- Citations
- 3
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 1.21