Reliability of Scheimpflug versus anterior segment optical coherence tomography after epithelial_off crosslinking in keratoconus

Background: corneal thickness monitoring is critical in the diagnosis, follow-up of keratoconic patients. This study aimed to evaluate the inter-device reliability in corneal pachymetry after epithelium-off crosslinking using anterior segment optical coherence tomography (AS-OCT) and Scheimpflug tomography. Method: This was a cross-sectional observational study conducted on patients who had epithelial-off crosslinking. Patients were subjected to follow-up imaging using Pentacam (Oculus) and AS-OCT (RTVue). The inter-device reliability was compared Results: 40 eyes, were included. Regarding central corneal thickness (CCT); pentacam measured a mean of 483.2 µm, versus 477.2 µm using AS-OCT. The thinnest corneal thickness (TCT) measured mean 470.05 µm with pentacam versus 444.7µm using AS-OCT, the mean paracentral superior corneal thickness (STC 2mm) by pentacam was 519.49µm versus 539.9µm by AS-OCT. Regarding mean paracentral inferior corneal thickness (ICT 2mm), pentacam measured 519.6µm versus 489.18µm by AS-OCT, the mean peripheral superior thickness by pentacam measured 589.6µm versus 565.43µm by AS-OCT. the mean peripheral inferior thickness (ICT 5mm), pentacam measured 578.7µm versus 529µm by AS-OCT . On evaluating the reliability between pentacam and AS-OCT by ICC it was excellent in (CCT), good in (TCT and STC 2mm), moderate in (STC 5mm), questionable in (ICT 2mm) and poor in (ICT 5mm). Conclusion: Based on the results, both AS-OCT and pentacam demonstrate comparable measurements with excellent reliability in the central corneal region, and good reliability in the thinnest location while this similarity decreased in the peripheral part, especially the inferior region.

Authors

Institutions

Publication Details

Journal
Benha Medical Journal
Published
2024-10-07
DOI
https://doi.org/10.21608/bmfj.2024.306976.2149
Primary Topic
Corneal surgery and disorders
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reliability of Scheimpflug versus anterior segment optical coherence tomography after epithelial_off crosslinking in keratoconus

Hamdy El-gazzar, Taher Kamel Eleiwa, Karima Mohamed Elgabry, Ahmed Sherin Mostafa
Benha Medical Journal
Corneal surgery and disorders
article

Reliability of Scheimpflug versus anterior segment optical coherence tomography after epithelial_off crosslinking in keratoconus

Hamdy El-gazzar, Taher Kamel Eleiwa, Karima Mohamed Elgabry, Ahmed Sherin Mostafa
article en

Abstract

Background: corneal thickness monitoring is critical in the diagnosis, follow-up of keratoconic patients. This study aimed to evaluate the inter-device reliability in corneal pachymetry after epithelium-off crosslinking using anterior segment optical coherence tomography (AS-OCT) and Scheimpflug tomography. Method: This was a cross-sectional observational study conducted on patients who had epithelial-off crosslinking. Patients were subjected to follow-up imaging using Pentacam (Oculus) and AS-OCT (RTVue). The inter-device reliability was compared Results: 40 eyes, were included. Regarding central corneal thickness (CCT); pentacam measured a mean of 483.2 µm, versus 477.2 µm using AS-OCT. The thinnest corneal thickness (TCT) measured mean 470.05 µm with pentacam versus 444.7µm using AS-OCT, the mean paracentral superior corneal thickness (STC 2mm) by pentacam was 519.49µm versus 539.9µm by AS-OCT. Regarding mean paracentral inferior corneal thickness (ICT 2mm), pentacam measured 519.6µm versus 489.18µm by AS-OCT, the mean peripheral superior thickness by pentacam measured 589.6µm versus 565.43µm by AS-OCT. the mean peripheral inferior thickness (ICT 5mm), pentacam measured 578.7µm versus 529µm by AS-OCT . On evaluating the reliability between pentacam and AS-OCT by ICC it was excellent in (CCT), good in (TCT and STC 2mm), moderate in (STC 5mm), questionable in (ICT 2mm) and poor in (ICT 5mm). Conclusion: Based on the results, both AS-OCT and pentacam demonstrate comparable measurements with excellent reliability in the central corneal region, and good reliability in the thinnest location while this similarity decreased in the peripheral part, especially the inferior region.

Benha Medical JournalVol. 0(0)
Mansoura University (EG), Benha University (EG)
Zero hunger
Openalex Percentile: Top 70%
Corneal surgery and disorders
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.