Tripod-Fix device for the treatment of painful osteoporotic vertebral compression fractures

Current vertebral augmentation procedures (VAPs) often involve devices associated with bone cement leakage. Tripod-Fix is designed to mitigate the risk of bone cement leakage by expanding in three dimensions to fit a narrower vertebral space. This study enrolled 12 patients diagnosed with osteoprorotic vertebral compression fractures (VCFs) for 12 month follow up. The primary outcomes assessed were changes in the Visual Analog Score (VAS) and Oswestry Disability Index (ODI) before and after treatment. Our results demonstrated significant pain relief with VAS decreasing from 8.58 ± 1.83 to 2.75 ± 1.54 cm and improved mobility with ODI decreasing from 73.67 ± 16.29 to 31.83 ± 23.33% post-treatment and sustained for 12 months. Follow-up radiographs revealed no device-related adverse events such as cement leakage, vertebral body collapse, or adjacent vertebral fractures (AVFs). In addition, the mean anterior height restoration ratio after treatment was 15.87 ± 5.13%. Our preliminary findings suggest that Tripod-Fix exhibits safety and efficacy comparable to the third-generation devices currently utilized for treating osteoporotic VCFs.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2024-10-12
DOI
https://doi.org/10.1038/s41598-024-75326-z
Citations
1
Primary Topic
Spinal Fractures and Fixation Techniques
Type
article
Field-Weighted Citation Impact
0.57
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Tripod-Fix device for the treatment of painful osteoporotic vertebral compression fractures

Shin-Yuan Chen, Hsiang‐Yi Hung
1 citations
Scientific Reports
Spinal Fractures and Fixation Techniques
0.57
article

Tripod-Fix device for the treatment of painful osteoporotic vertebral compression fractures

Shin-Yuan Chen, Hsiang‐Yi Hung
article en
1 citations

Abstract

Current vertebral augmentation procedures (VAPs) often involve devices associated with bone cement leakage. Tripod-Fix is designed to mitigate the risk of bone cement leakage by expanding in three dimensions to fit a narrower vertebral space. This study enrolled 12 patients diagnosed with osteoprorotic vertebral compression fractures (VCFs) for 12 month follow up. The primary outcomes assessed were changes in the Visual Analog Score (VAS) and Oswestry Disability Index (ODI) before and after treatment. Our results demonstrated significant pain relief with VAS decreasing from 8.58 ± 1.83 to 2.75 ± 1.54 cm and improved mobility with ODI decreasing from 73.67 ± 16.29 to 31.83 ± 23.33% post-treatment and sustained for 12 months. Follow-up radiographs revealed no device-related adverse events such as cement leakage, vertebral body collapse, or adjacent vertebral fractures (AVFs). In addition, the mean anterior height restoration ratio after treatment was 15.87 ± 5.13%. Our preliminary findings suggest that Tripod-Fix exhibits safety and efficacy comparable to the third-generation devices currently utilized for treating osteoporotic VCFs.

Scientific ReportsVol. 14(1)
Tzu Chi University (TW), Hualien Tzu Chi Medical Center (TW)
Openalex Percentile: Top 29%
Spinal Fractures and Fixation Techniques
0.57
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Tripod-Fix device for the treatment of painful osteoporotic vertebral compression fractures — Shin-Yuan Chen, Hsiang‐Yi Hung · Scientific Reports (2024) | TGRS Research Map | TGRS