The out-of-plane contact shield and mA-modulation – the effect on fetal absorbed dose

The effect of patient shielding on fetal radiation dose was evaluated in computed tomography pulmonary angiography with the out-of-plane shield visible in the localizer but absent in the scan range in chest computed tomography (CT). An anthropomorphic phantom with additional prosthetic pregnancy belly was scanned with different CT scanners using clinical imaging protocols and radiophotoluminescence dosemeters (GD-352 M). The out-of-plane shield decreased the fetal absorbed radiation dose with Siemens Somatom go.Up, Canon Aquilion Prime SP and Canon Aquilion One scanners. The decrease was 3.9%-39.4% (0.01-0.09 mGy). With GE Optima the shield increased the fetal dose by 100% (0.23 mGy), with Canon Aquilion One and GE Optima scanners the abdomen dose increase was 17.5% and 36.4%, respectively (0.61 and 1.38 mGy). Applying an out-of-plane shield outside the scanned volume may increase the fetal radiation dose during CT when using tube current modulation, depending on the make and model of the CT scanner.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2024-10-11
DOI
https://doi.org/10.1093/rpd/ncae204
Citations
2
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.81

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The out-of-plane contact shield and mA-modulation – the effect on fetal absorbed dose

Riitta Parkkola, Chibuzor T. M. Eneh, Heli R S Larjava, Aleksi Saikkonen
2 citations
Radiation Protection Dosimetry
Radiation Dose and Imaging
0.81
article

The out-of-plane contact shield and mA-modulation – the effect on fetal absorbed dose

Riitta Parkkola, Chibuzor T. M. Eneh, Heli R S Larjava, Aleksi Saikkonen
article en
2 citations

Abstract

The effect of patient shielding on fetal radiation dose was evaluated in computed tomography pulmonary angiography with the out-of-plane shield visible in the localizer but absent in the scan range in chest computed tomography (CT). An anthropomorphic phantom with additional prosthetic pregnancy belly was scanned with different CT scanners using clinical imaging protocols and radiophotoluminescence dosemeters (GD-352 M). The out-of-plane shield decreased the fetal absorbed radiation dose with Siemens Somatom go.Up, Canon Aquilion Prime SP and Canon Aquilion One scanners. The decrease was 3.9%-39.4% (0.01-0.09 mGy). With GE Optima the shield increased the fetal dose by 100% (0.23 mGy), with Canon Aquilion One and GE Optima scanners the abdomen dose increase was 17.5% and 36.4%, respectively (0.61 and 1.38 mGy). Applying an out-of-plane shield outside the scanned volume may increase the fetal radiation dose during CT when using tube current modulation, depending on the make and model of the CT scanner.

Radiation Protection DosimetryVol. 200(19)
University of Turku (FI), Turku University Hospital (FI)
Turun Yliopistollinen Keskussairaala
Good health and well-being
Openalex Percentile: Top 25%
Radiation Dose and Imaging
0.81
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.