Utilizing Low Sample Uptake Rates and A Nitrogen Mixed-gas Plasma for the Elimination of Oxide-based Interferences in ICPMS Analyses

Spectroscopic interferences have long negatively impacted the accuracy of inductively coupled plasma mass spectrometry (ICPMS) analyses.Of these, oxide-based interferences, the combination of an analyte with oxygen producing a new ion 16 mass units greater than the original analyte, often proves most prevalent and difficult.A cheap and reliable method that permits the mitigation of oxide-based interference would be highly beneficial.Here-in, low sample uptake rate was used to reduce the formation of lanthanide oxide-based interferences in ICPMS analyses through temperature and Le Chtelier effects.Introduction of oxide forming solutions (50 g L -1 ) composed of lanthanide elements at 1 mL/min yielded an average oxide ratio of 4.5 7.2% while introduction at 50 L L min -1 yielded 0.54 0.26%.A similar method using 2% nitrogen gas in the bulk plasma concurrently decreased oxide-based interferences.The benefits observed with low sample uptake rate and a mixed-gas plasma were combined to virtually eliminate oxide based-interferences for many of the lanthanide elements and provide a modest signal enhancement compared to an Ar plasma operated at a higher sample uptake rate.For example, when comparing the best oxide reduction method to the worst, oxide formation is mitigated by 97%.Of the three sample uptake rates tested, 235 L min -1 under mixed-gas plasma conditions offers the best balance between the oxide interferences mitigation and signal intensity.Ultimately, low sample uptake rate may prove essential in increasing ICPMS analysis accuracy while safeguarding resources and minimizing chemical waste for generations to come.

Authors

Institutions

Publication Details

Journal
Atomic Spectroscopy
Published
2024-08-31
DOI
https://doi.org/10.46770/as.2024.174
Citations
2
Primary Topic
Analytical chemistry methods development
Type
article
Field-Weighted Citation Impact
0.47

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Utilizing Low Sample Uptake Rates and A Nitrogen Mixed-gas Plasma for the Elimination of Oxide-based Interferences in ICPMS Analyses

Diane Beauchemin
2 citations
Atomic Spectroscopy
Analytical chemistry methods development
0.47
article

Utilizing Low Sample Uptake Rates and A Nitrogen Mixed-gas Plasma for the Elimination of Oxide-based Interferences in ICPMS Analyses

Diane Beauchemin
article en
2 citations

Abstract

Spectroscopic interferences have long negatively impacted the accuracy of inductively coupled plasma mass spectrometry (ICPMS) analyses.Of these, oxide-based interferences, the combination of an analyte with oxygen producing a new ion 16 mass units greater than the original analyte, often proves most prevalent and difficult.A cheap and reliable method that permits the mitigation of oxide-based interference would be highly beneficial.Here-in, low sample uptake rate was used to reduce the formation of lanthanide oxide-based interferences in ICPMS analyses through temperature and Le Chtelier effects.Introduction of oxide forming solutions (50 g L -1 ) composed of lanthanide elements at 1 mL/min yielded an average oxide ratio of 4.5 7.2% while introduction at 50 L L min -1 yielded 0.54 0.26%.A similar method using 2% nitrogen gas in the bulk plasma concurrently decreased oxide-based interferences.The benefits observed with low sample uptake rate and a mixed-gas plasma were combined to virtually eliminate oxide based-interferences for many of the lanthanide elements and provide a modest signal enhancement compared to an Ar plasma operated at a higher sample uptake rate.For example, when comparing the best oxide reduction method to the worst, oxide formation is mitigated by 97%.Of the three sample uptake rates tested, 235 L min -1 under mixed-gas plasma conditions offers the best balance between the oxide interferences mitigation and signal intensity.Ultimately, low sample uptake rate may prove essential in increasing ICPMS analysis accuracy while safeguarding resources and minimizing chemical waste for generations to come.

Atomic SpectroscopyVol. 45(04)
Queen's University (CA)
Queen's University, Natural Sciences and Engineering Research Council of Canada
Clean water and sanitation
Openalex Percentile: Top 41%
Analytical chemistry methods development
0.47
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.